1 9
SYSTEM
E N
INTELECT
®
MOBILE 2 ULTRASOUND USER MANUAL
POWER
Input
100 - 240 V AC, 1.0 to 0.42 A, 50/60 Hz
Electrical Class
CLASS II
Mode of Operation
Continuous
Note: Mains isolation is achieved by use of the double pole switch
located on the rear panel.
Electrical Type (Degree of Protection)
Ultrasound
.TYPE BF
SYSTEM SPECIFICATIONS AND DIMENSIONS
Width
Depth
Height
Weight (no battery)
Intelect Mobile 2 Head Unit
UltraSound
34cm
35.5cm
15cm
2.8kg
Cart configurations
Cart (Safe working load 6.5kg)
48cm (MAX)
52cm (MAX)
96cm
10.1kg
Device on cart
-
-
111 cm
-
ULTRASOUND SPECIFICATIONS
Frequency
1 MHz; 3 MHz
Duty Cycles
10%, 20%, 50%, Continuous
Pulse Repetition Rate
16, 48, or 100 Hz
Pulse duration:
1 -31.25 ms
Max (ON): 31.25 ms
Min (OFF): 5ms
OUTPUT POWER
US applicator
Frequency
1 cm
2
2cm
2
5cm
2
1MHz
3MHz
1MHz
3MHz
1MHz
3MHz
Effective Radiating
Area ERA INTL (cm2)
1
0.9
1.5
1
2.5
2.7
Max Output power
in Continuous mode
2W
1.8W
3W
2W
5W
5.4W
Max Output power
in Pulsed mode
3W
2.7W
(*)
4.5W
3W
7.5W
8.1W
Max Amplitude in
Continuous mode
2W/
cm
2
2W/
cm
2
2W/
cm
2
2W/
cm
2
2W/
cm
2
2W/
cm
2
Max Amplitude in
Pulsed mode
3W/
cm
2
3W/
cm
2
3W/
cm
2
3W/
cm
2
3W/
cm
2
3W/
cm
2
(*) An error of + 0.25 W can be measured with 1cm2 US applicator,
pulse mode 100Hz at 10% or 20% Duty Cycle.