MOBILE 2 ULTRASOUND USER MANUAL
Unless otherwise specified, ultrasound
controls accuracy is:
± 20 %.
Peak to Average Ratio:
1:1, at 50% Duty Cycle
4:1, at 20% Duty Cycle
9:1, at 10% Duty Cycle
Beam Nonuniformity Ratio
<5:1
Beam Type
Collimating
Treatment Time
1 to 30 min
GENERAL SYSTEM OPERATING AND
STORAGE TEMPERATURE
Operating Conditions
The device will meet its requirement under the following
conditions:
Temperature:
5°C to 40°C
Relative Humidity:
15% to 90%
Atmospheric Pressure:
70kPa to 106kPa
Transport and Storage Conditions
The device will remain in proper condition under the
following conditions:
Temperature:
-20°C to 60°C
Relative Humidity:
10% to 90%
Atmospheric Pressure:
50kPa to 106kPa
Time required for the Intelect Mobile 2 to warm from the
minimum storage temperature between uses until the
Intelect Mobile 2 is ready for its INTENDED USE when the
ambient temperature is 20 °C: 5h
Time required for the Intelect Mobile 2 to cool from the
maximum storage temperature between uses until the
Intelect Mobile 2 is ready for its INTENDED USE when the
ambient temperature is 20 °C: 5h
IPXX Rating for Unit
Rated to IP21
IP2* Protection against fingers or other object not greater
than 80mm in length and 12mm in diameter
*1 Protection from vertically dripping water
IPXX Rating for US applicator
Rated to IPX7
IPX7 Protection from immersed in water (up to 1m depth)
RED
RF transmitter/receiver characteristics:
- Frequency Band transmission:
2400–2483.5 MHz
- Modulation type:
GFSK
- Data rate:
up to 2Mbps 500kHz deviation at 2Mbps
- Effective radiated power:
+6dBm