7
Precautionary Instructions
1.
CAUTION:
Read, understand and practice the precautionary and operating instructions. Know the limitations
associated with the Hydrocollator Master Heating Units. Observe the cautionary and operational
decals installed on the unit.
2.
CAUTION:
The thermostat is extremely sensitive and the slightest adjustment will alter the temperature several
degrees. The recommended operating temperature is 160° F to 166° F (71° to 74°C). The temperature
of the water should be checked with a thermometer after every adjustment, before using the
HotPacs. Always allow sufficient time for the water temperature to stabilize.
3.
CAUTION:
ALWAYS keep the water level over the top of the HotPacs to avoid burning or scorching of the packs.
4.
CAUTION:
NEVER allow the HotPacs to dry after use.
5.
CAUTION:
Check water level daily as it has a natural loss due to evaporation.
6.
CAUTION:
NEVER adjust the thermostat too high. The thermostat is extremely sensitive to the slightest
adjustment.
7.
CAUTION:
Clean the tank periodically as described in the maintenance portion of this manual.
8.
CAUTION:
ALWAYS replace Hydrocollator HotPacs as soon as they show signs of wear.
9.
CAUTION:
If the unit is to be left unattended for a period of time, unplug, remove the packs, empty the water
and clean the tank.
10.
CAUTION:
ALWAYS use factory authorized replacement parts.
11.
CAUTION:
ALWAYS destroy and replace HotPacs which are exposed to contamination.
12.
CAUTION:
Certain additives (such as herbal teas, essential oils, etc.) will damage the heating element,
thermostat and stainless steel of the heating unit. The use of such additives voids the manufacturer's
warranty.
13.
NOTICE:
Install Hotpacs in rack before filling the tank with water to ensure proper operation of your heating unit
as well as correct water level. If using distilled water, it is imperative that the HotPacs are placed in
the unit first.
14.
NOTICE:
Chlorine in regular tap water may be present in high enough concentrations to damage your unit by
causing rust. We highly recommend that you clean your unit regularly. If you suspect high levels of
chlorine in your Hydrocollator Heating Unit, we recommend the addition of a dechlorinator
15.
WARNING:
Grounding the Hydrocollator Master Heating Unit: Make certain the unit is electrically grounded by
plugging into a hospital grade electrical outlet with a ground terminal receptacle (U-ground outlet, 120
VAC only). Follow the National Electrical Code.
16.
WARNING:
The water temperature in the Hydrocollator is approximately 160° F (71° C), and the water scalding
temperature is approximately 120° F (49° C).
General Operating Instructions
The Hydrocollator is simple and easy to operate. The controls are located in the rear, near the bottom of the unit. The
thermostat controls the temperature of the water. It is located near the bottom of the unit on the control panel. It has
been tested and preset at the factory to normal operating temperature. If further adjustment is necessary, it can be
easily accomplished by turning the adjusting screw located on the thermostat. Turning the outer screw
counterclockwise lowers the temperature, while turning it clockwise raises it.
NOTE:
Do not turn the inner stem.
The Hydrocollator units operate through the use of a hydraulic capillary-type thermostat and can be left plugged in and
turned "ON" to provide a ready supply of heated HotPacs.
The HotPac™ rack is adjustable and accommodates different sizes and types of packs.
SS-2 Unit – Adjustment is accomplished by placing the rack insert in or leaving the insert out. The unit comes complete
with eight standard HotPacs.
M-2 and M-4 Units – Adjustment is accomplished by inserting the hanging platform. The M-2 unit comes complete
with 12 standard HotPacs. The M-4 comes complete with 24 standard HotPacs.