- 19 -
Remove full scale
load for zero
verification
T
PEAK to go back
LBF
0.0
INFO to proceed
Verify tension
calibration
T
PEAK to go back
LBF
100.0
INFO to proceed
Save new calibration
T
PEAK to go back
LBF
0.0
INFO to proceed
C
Note: Remove all weights but leave tension fixture at-
tached to the force gauge.
C
C
Note: Verify the tension calibration by checking both the
full scale (span) and zero. When no load is applied, the
display should read zero. When full scale weights are
applied to the fixture, the display should read the full
scale weight equivalent.
Note: Select INFO to save the calibration.
q
q
i
i
i
www.
GlobalTestSupply
.com
Find Quality Products Online at: