- 17 -
Procedure
The Chatillon DFX II Series is supplied with a Cer-
tificate of Conformance. The gauge was tested and
found to have a measurement accuracy of 0.5% of
full scale or better.
Calibration Procedure
The DFX II Series calibration procedure is contained
in the gauge firmware. The calibration is a four
point calibration: full scale compression, zero com-
pression, full scale tension, zero tension.
To access the calibration routine, begin with the
gauge power OFF.
Press the ON key. While the “Chatillon Logo” is
displayed, press PEAK then UNITS.
Model DFX II-100
Peak
Rev. No. V1.00
Chatillon
Rev. Date
dd/mm/yy
T
C
Units
Gauge Service
Peak
UNITS to diagnostics
T
C
PEAK to calibrate
INFO to exit
Select Gauge Model
Peak
UNITS to change
T
C
INFO to proceed
PEAK to exit
Select CALIB Units
i
UNITS to change
INFO to proceed
T
C
DFX II-100
PEAK to go back
LBF
0.0
Note: The units of measure selected must correspond to
the weights that are being used for the calibration.
Place gauge in
i
horizontal position
T
C
PEAK to go back
LBF
0.0
INFO to proceed
Note: Placing the gauge horizontally means to lay the
gauge perfectly flat on its backside making sure no load
is being applied to the loadcell shaft.
Place gauge in
i
compression mode and
exercise 3 times
T
C
PEAK to go back
LBF
0.0
INFO to proceed
Note: You must use a hanger when performing the
compression calibration routine to eliminate side load
effects on the loadcell.
Note
Selecting PEAK during the calibration procedure
will take you back to the last step in the procedure.