20994124 REV A
20
Fusion electronics; in addition to the size requirement, the probe must be a
shielded two-wire (+ and – signal wires, and shield), “PT1000” type, having a
nominal resistance of 1000 ohm at 0
°
C. For calibration procedure, see section
7.3.
Figure 12: Fusion temperature probes. (a) Temperature probe on the back panel
of the liquefier, (b) Additional temperature probe access tube for the secondary
temperature probe
3.7 Product Specifications
Table 2 : MVE Fusion Product Specifications
Safety for Electrical Equipment
and Lab Use
ETL marked for use in the United States and Canada
according to:
UL 61010-1:2012 (Edition 3.0)
CSA/CAN 22.2 #61010-1:2012 (Edition 3.0)
ASME Section VIII (Cryogen pressure vessel)
Physical Dimensions
Height: < 65 inches (1625 mm)
Width: 43.5 inches (1088 mm)
Depth: 57 inches (1450 mm)