Lid and Cork
- The lid covering the storage space (shown in
Figure
8
Error! Reference source not found.
), the gasket on the rim of the dewar
neck, and the gasket underneath the lip of the lid are all replaceable parts.
Together, they provide insulation and a nearly airtight seal to prevent heat and
moisture from entering the storage space.
Retractable Stairs
- The Built-in stairs make it easy to access the storage space.
They can be folded up against the freezer when not in use, as shown in
Figure
8
. A restraint is included to keep the stairs from being accidentally unfolded.
Transfer line –
In the event the MVE Fusion needs refilling or topping off (for
instance after a prolonged power outage) a proper hose to transfer LN2 from a
supply tank to the pressure vessel is required. See “filling the MVE Fusion” for
more details. A vacuum-jacketed hose with ½” flare fittings on each is required;
Chart recommends using a transfer hose with a “candy cane” pressure-relief