2-6 SECTION 2
–
OPERATING INSTRUCTIONS
2.
Open the DRAIN VALVE (#20) and AIR-TO-CYLINDER VALVE (#22). After the
fluid has been drained from the cylinder, close the AIR TO CYLINDER VALVE.
3.
Loosen the gland on the center thermocouple in cylinder plug, bleed off all the air and
remove the thermocouple.
4.
Remove the cylinder plug.
Water Flushing the Cylinder
1.
Turn “ON” the AIR EXHAUST VALVE SWITCH (#
9). Wait for the pressure to
decrease to zero.
2.
Open the PRESSURE RELEASE VALVE (#21) and FLUSH WATER VALVE (#19).
3.
When the cylinder has been sufficiently flushed, close the FLUSH WATER VALVE
(#19) and PRESSURE RELEASE VALVE (#21).
4.
Open the DRAIN VALVE (#20) and AIR-TO-CYLINDER VALVE (#22). After the
fluid has been drained from the cylinder, close the AIR TO CYLINDER VALVE.
5.
Loosen the gland on the center thermocouple in cylinder plug, bleed off all the air and
remove the thermocouple.
6.
Remove the cylinder plug.