Section 4 - Calibration
Note: The pressure transducers, RTDs, and flow meters were all calibrated by their
respective manufacturers. See their calibration sheets in the accompanying
document package. The CHA in
strument’s measurements of pressures
,
temperatures, and flow rates (using those transducers, RTDs, and flow meters)
were all calibrated at Chandler Engineering.
Pressure Calibration
The three pressure transducers used by the Model 7200 CHA are very-high-accuracy (+/-
0.1% of Full Scale) transducers. These transducers output a linear 0-5VDC signal
corresponding to 0-1000 psig. This signal is read by the data-acquisition module and the
CHA-Control software converts this voltage to a pressure value according to the calibration
scaling entered on the calibration screen for that transducer. The procedure is identical for all
three pressures.
Figure 13 - Pressure Calibration
There are three methods of calibration:
1.
Auto Calibrate (Recommended):
Let the software use the Shunt Calibration feature
of the transducer to automatically calibrate it and scale the conversion from voltage to
pressure.
2.
Span and Zero Calibration:
Apply exactly known pressures to the transducer and
the software will calculate the Slope and Intercept.
3.
Manual Entry:
Manually enter a Slope and Intercept.
Summary of Contents for AMETEK 7200
Page 10: ......
Page 16: ...2 2 SECTION 2 OPERATING INSTRUCTIONS Figure 1 Front Panel...
Page 38: ......
Page 48: ...5 2 SECTION 5 TROUBLESHOOTING GUIDE This page is intentionally left blank...
Page 50: ...6 2 SECTION 6 REPLACEMENT PARTS This page is intentionally left blank...
Page 52: ...6 2 SECTION 6 REPLACEMENT PARTS This page is intentionally left blank...
Page 65: ...CHANDLER ENGINEERING...
Page 66: ......