Page: 17/21 User Manual PULS-CTRL24, Rev.1-00: Software Utilities
100495 (1888Fh) bytes processed, 99584 (18500h)
bytes programmed
Resetting the target
Program finished ok
During the programming procedure, a message box is displayed at
the device display. When the programming finishes, the device is re-
started with the new firmware.
If an error occurs, the flash loader utility at the microcontroller may
remain active. This is the case if the message box at the device dis-
play is still present and the device did not restart. In this case, you
may retry the action with the command line parameter '-i':
FlashLoader 6 Firmware.txt -i
This will prevent the utility at the host computer from initializing the
flash loader utility at the microcontroller again and it will just try to re-
program the file
Firmware.txt
. If the error persists, contact the
manufacturer.
!
Attention:
You must not power down the device if the firmware up-
grade did not succeed. Otherwise, the device will not operate properly
or might even not restart at all. Did this happen, it would be necessary
to reprogram the device in the factory.
If the current firmware is damaged so that the device is inoperable,
you may try to start the flash loader utility at the microcontroller manu-
ally. Press, simultaneously, the horizontal direction keys (left and
right) and power on the device. If this small part of the firmware is still
working, the device's flash loader will start. Then, try to launch the util-
ity FlashLoader with the command line parameter '-s':
FlashLoader 6 Firmware.txt -s
This does not start the flash loader utility at the microcontroller but will
only try to reprogram the file
Firmware.txt
. If an error occurs that
you cannot solve, contact the manufacturer.
CGC Instruments
Tel.: +49 (371) 355 098–55
Hübschmannstr. 18 | D–09112 Chemnitz
Fax: +49 (371) 355 098–60
internet: www.cgc-instruments.com
e–mail: [email protected]