Page: 34/81 User Manual PSU-CTRL-2D, Rev.1-00: Software Utilities
Tab. 9. Command line parameters of the program
COM-HVPSU2D-
Control
- Backup and restore.
Parameter Explanation
-y
FileName
backup the NVM data into a text file with
the name
FileName
-Y
FileName
restore the NVM data from a text file with
the name
FileName
Backing Up and Restoring the Data
All data stored in the device's NVM, i.e. the device settings including
the current configuration as well as the user configurations can be
backed up or restored (see Tab. 9). A restore procedure rolls back the
device exactly to the state it was in at the time of the backup. Thus, if
any adjusting of the settings is planned, it is recommended to first
create a backup with which the original state can be restored.
Since the back up and restore procedures transfer a large amount of
data, it is recommended to increase the communication speed by
specifying the parameter
-$
(see section "General Control"). To back
up the system memory into a data file
Memory.txt
at a communica-
tion speed of 230.4 kbaud, execute the following command:
COM-HVPSU2D-Control 6 -$230400 -y MemoryData.txt
-t
This command downloads the memory data from the device into the
file
Memory.txt
.
To restore the data at the same communication speed, execute the
following command:
COM-HVPSU2D-Control 6 -$230400 -Y MemoryData.txt
-t
This command uploads the memory data from the file
Memory.txt
to
the device. Since all data in the NVM will be overwritten, the utility
asks for a confirmation twice.
CGC Instruments
Tel.: +49 (371) 355 098–55
Hübschmannstr. 18 | D–09112 Chemnitz
Fax: +49 (371) 355 098–60
internet: www.cgc-instruments.com
e–mail: [email protected]