3
X/Y camera too bright or too
dark
Incorrect setting of CMOS gain
Reset the gain value
Abnormal display of X/Y
camera
1. Camera
plate
cable
not
plugged
2. Damage of camera plate or
control board
1. Plug in the camera plate
cable
2. Replace the camera plate or
control board
Left/right motor error at start
1. Propel motor cable not plugged
2. Foreign
material
between
propel motor gears
3. Jam of propel motor
1. Plug in the propel motor cable
2. Clean up the foreign objects
3. Replace the propel motor
Abruption
or
abnormal
noise of propel motor
1. Foreign objects between the
propel motor gears
2. Propel motor fault
1. Clean up the foreign objects
2. Replace the propel motor
"Displace motor error X/Y"
at aligning
1. Align (displace) motor cable not
plugged
2. Jam of align (displace) motor
1. Plug in the align (displace)
motor cable
2. Replace the align (displace)
motor
No discharge
1. High-voltage discharge plate
cable not plugged
2. Damage
of
high-voltage
discharge plate
1. Plug
in
the
high-voltage
discharge plate cable
2. Replace
the
high-voltage
discharge plate
Discharge to V-groove
Offset of electrode position
Adjust the electrode position
Weak or unstable discharge
1. Improper setting of discharge
parameters
2. Dirty electrode
1. Adjust the discharge parameters
or perform the discharge correction
2. Clean or replace the electrode
Heater failure
1. Heater cable not plugged
2. Heater damage
1. Plug in the heater cable
2. Replace the heater
Auto heating failure if the
heater cover is closed
1. Manual operation of heater
2. Damage
of
heater
optical
switch
1. Set it to auto
2. Replace the heater
Fiber not held by heater
cover
Magnetic steel shedding
1. Re-bind the magnetic steel
2. Replace the heater
Fiber not held by fixture
pressure plate
1. Fixture
magnetic
steel
shedding
2. Upper and lower sliders not
came out
1. Re-bind the magnetic steel
2. Replace the fixture
Obvious tilt of X/Y fiber
image
Minor changes in camera plate
position
Re-adjust the camera plate
Dark spots or shadows on
X/Y video background
Dirt on the X/Y microscope lens or
CMOS surface
Clean the lens or CMOS
3.2 Causes of Fusion Loss Increase and Solutions