4957D/E/F Microwave Analyzer Quick Start Guide 

parameters  such  as  reference,  scale,  max  and  min  values,  relative  measurement 
switch  state,  offset,  etc..  In  the  Power  Monitor  mode,  it  can  also  turns  on  /  off  the 
preamplifier. 

  [BW]: It is used to set such parameters as Average, Smooth and MF BW, etc., in the 

CAT mode and the Network Analyzer mode. In the Spectrum Analyzer mode and the 
Power Monitor, it is used to set RBW, VBW, Detector and Average, etc.. 

  [Mkr]:  In  the  CAT  mode  and  Network  Analyzer  mode,  it  is  used  to  open  function 

menus  related  to  markers,  including  marker  switching,  opening  and  closing,  Mkr 
Peak  Search,  Delta  Mkr  and  marker  function  settings.  In  the  Spectrum  Analyzer 
mode,  it  is  used  to  open  function  menus  related  to  markers,  including  marker 
switching, opening and closing, Mkr Peak Search, Delta Mkr, Counter and Noise Mkr 
settings. 

 

Measure

: In the CAT mode, it is used to select the data format of measurement, 

including  VSWR,  Return  Loss,  Impedance,  Cable  Loss,  DTF  VSWR  and 
DTF@Return Loss. In the Network Analyzer mode, it is used to select the required S 
parameter measurement mode, data format and time domain function settings. In the 
Spectrum  Analysis  mode,  it  is  used  to  select  different  measurement  functions, 
including field strength measurement, channel power, OBW, AM / FM DM, ACPR, etc. 
or  set  signal  source  power  output.  In  the  Vector  Voltmeter  mode,  it  is  used  to  set 
such parameters as measurement type, port selection, measurement format, relative 
measurement switch and output power, etc.. 

  [Mode]: It is used to select the working mode of the analyzer, which can be selected 

as  CAT,  Network  Analyzer,  Spectrum  Analyzer,  Power  Monitor,  USB  Power  Meter 
and Voltage Voltmeter, etc.. 

  [Cal]:  It  is  used  to  select  the  type  of  device  under  calibration  piece,  perform 

calibration,  turn  calibration  on  and  off,  and  resume  calibration  in  the  CAT  mode, 
Network Analyzer mode, and Vector Voltmeter mode. 

  [Trace]:  It  is  used  to  set  such  parameters  as  trace  calculation  and  limit  line  in  the 

CAT  mode  and  the  Network  Analyzer  mode.  In  the  Spectrum  Analyzer  mode,  it  is 
used for  switching,  opening,  closing,  holding  and  refreshing  traces and  setting  limit 
lines, etc.. 

  [Sweep]:  It  is  used  to  set  such  parameters  as  Swp  Time,  Points,  Swp  mode  and 

Trigger  mode  in  the  CAT  mode  and  the  Network  Analyzer  mode.  In  the  Spectrum 
Analyzer mode, it is used to set such parameters as Swp Time, Swp Type and Swp 
Mode, etc.. In the Power Monitor mode, it is used to set Swp Type. 

  [System]:  In  the  local  working  mode,  it  is  used  for  system  date,  time  and  power 

saving  mode,  viewing  electronic  serial  numbers  of  products,  selecting  system 
operation  language  and  other  system-related  settings;  When  the  analyzer  is  under 
remote control, press this key to restore local functions. 

  [File]:  It  is  used  for  setting  the  working  state  of  the  analyzer  as  well  as  storing  and 

calling measurement data. 

 

[↑]  and  [↓]:  Are  used  to  control  stepping  up and  down  or  to  move  up and down  to 
select the current item. 

 

[←] Backspace key: Is used to delete the last number or character entered. 

  [Cancel]:  Is  used  to  ignore  the  settings  and  input  in  the  dialog  box  and  close  the 

dialog box. 

  [OK]: Is used to confirm the settings and input values in the dialog box and close the 

dialog box. 

  [Reset]:  Is  used  for  system  reset,  system  restart  and  restoring  to  the  default  initial 

state. Press this key and release it to reset. 

Summary of Contents for 4957D

Page 1: ...4957D E F Microwave Analyzer Quick Start Guide China Electronics Technology Instruments Co Ltd...

Page 2: ......

Page 3: ...duct can meet the indicator requirement in the manual upon delivery The calibration and measurement are completed by measuring bodies with national qualification with relevant data to be provided for...

Page 4: ...A Mode 9 3 2 DTF Measurement in the CAT Mode 15 3 3 Signal Measurement in the SA Mode 16 3 4 Measuring Signal Power in the Power Monitor PM Mode 18 3 5 Measuring the Electrical Length of Cables or Oth...

Page 5: ...the following chapters and sections Get Prepared This chapter introduces the preparation before operation startup panel introduction and battery replacement of 4957D E F After reading this chapter you...

Page 6: ...the instrument not mentioned in the manual Otherwise such consequences as reduced electromagnetic shielding property and internal component damage can occur affecting product reliability In this case...

Page 7: ...e The basic parameters of the enclosed battery are as follows Nominal voltage 10 8 V Nominal capacity 7 800 mAh Working time About 2 5 h Charging time About 6 h Rechargeable batteries should not be ex...

Page 8: ...s on the bottom of the touch screen are represented by XXX with XXX being the name of the key The menu keys on the right are represented by XXX with XXX being the menu name 2 3 Front Panel This sectio...

Page 9: ...ch can be selected as CAT Network Analyzer Spectrum Analyzer Power Monitor USB Power Meter and Voltage Voltmeter etc Cal It is used to select the type of device under calibration piece perform calibra...

Page 10: ...installed power connected b Battery installed and full power connected Flashing yellow a Battery installed and not full power connected Power on Normal on green a Battery not installed power connected...

Page 11: ...arked as 10 MHz In Out with 50 impedance BNC female adapter 4957D E F microwave analyzers can receive externally input 10 MHz sine signals as phase locked reference signals and the input signals shoul...

Page 12: ...ally switched from the speaker to the earphone 2 5 4 Instrument Symbols The instrument symbol yellow label shown in the figure indicates that the max power input from the test spectrum input port and...

Page 13: ...g soft key to select the correct calibration kit model 5 Press OK return to the menu bar of Cal select the frequency response error correction function press M Cal Freq Resp OPEN S11 and the system wi...

Page 14: ...pt Connect Thru cable between P1 and P2 then press corresp menu button to start Then connect N type cable between ports 1 and 2 All adapters or cables used in device measurement should be included i e...

Page 15: ...location where the DUT is to be connected 5 After the display trace is stable press THRU and the analyzer will display Cal std measuring during the measurement of the standard piece and after completi...

Page 16: ...etc 3 Open the calibration menu and press Cal 4 Press Cal Cal Kit to see if the calibration kit is consistent with the kit specified with the current soft key If not press the corresponding soft key t...

Page 17: ...the breaker and connect the load to the test port 10 After the display trace is stable to measure the standard piece press LOAD the analyzer will display calibration standard measurement during the me...

Page 18: ...e and connect the short circuit device to the test port 1 After the display trace is stable to measure the standard piece press P1 SHORT the analyzer will display calibration standard measurement duri...

Page 19: ...layed in the information zone Connect the DUT for measure Figure 3 5 Standard piece connection of full 2 ports error correction In actual calibration there is no sequence for the open circuit device s...

Page 20: ...w to display the submenu of Window and then select different window functions 5 Press Back Unit m ft to set the required distance unit system 6 Press Cable Specification and Load Cable to select the m...

Page 21: ...ency value 4 Setting sweep span Press Freq Span Note whether the span data is displayed in the information input zone to determine the currently activated parameters Reduce the span to 20MHz for insta...

Page 22: ...eed of connection to an external power probe can conveniently test the power of continuous wave signals with a certain bandwidth and the channel bandwidth can be set to a larger value so as to simulat...

Page 23: ...onvenient operation interface as the vector voltmeter The results of single port electrical length measurement are displayed in amplitude and phase formats while the results of all port are only displ...

Page 24: ...port error correction in the VVM mode 5 Selecting measurement mode The VVM mode of the analyzer allows for measurement of parameters of the DUT like amplitude phase VSWR and impedance etc 1 Press Mea...

Page 25: ...USB Power Meter mode and press OK to complete Figure 3 11 Test interface of USB Power Meter mode 2 Connecting power probes 1 As shown in the figure connect the small end of the USB cable to 8723X ser...

Page 26: ...required to set the frequency value of the measured CW Freq signal The USB power probe will store a set of correction factors for each frequency The frequency corresponding to the measured power signa...

Page 27: ...e losses for added attenuators or connecting cables It also allows for setting power offsets for gains of added amplifiers Positive values can compensate losses and negative values can compensate gain...

Page 28: ...F microwave analyzer provides two limit lines upper and lower When the amplitude of a signal in a certain frequency band exceeds the set upper limit or is less than the set lower limit the analyzer w...

Page 29: ...rmat or csv data text format The Print Screen function allows for saving the information displayed on the current screen into the memory of the instrument in an image format The storage medium for sta...

Page 30: ...environment for compliance with the requirements the test port connectors for any damages and the calibration piece for normal performance indicators If everything above is normal the instrument may b...

Page 31: ...g to the following steps 1 Write a paper document describing the instrument failure and put it in the package together with the analyzer 2 Pack it with the original packing materials so as to minimize...

Page 32: ...y 32 dB 40 dB 32 dB 40 dB 28 dB 35 dB Effective source matching 25 dB 37 dB 25 dB 37 dB 18 dB 30 dB Reflected tracking 0 10dB 0 14dB 0 10dB 0 25dB 0 10dB 0 25dB Transmission tracking 0 25dB 0 33dB 0 2...

Page 33: ...nput level 27 dBm Test port type N type female 3 5 mm male 2 4 mm male Power supply Rechargeable li ion battery or power adapter Power consumption 45 W excluding battery charging Working temperature r...

Page 34: ...a test software 4957D S03 Vector voltmeter software 4957D S04 USB power measurement software USB power probe to be equipped separately 4957D S05 Power meter software 4957D S06 Field strength measureme...

Page 35: ...female male calibration cable GORE 0K0CJ0CK024 0 4957D H14 87230 USB CW power probe 4957D H15 87231 USB CW power probe 4957D H16 87232 USB CW power probe 4957D H17 87233 USB CW power probe 4957D H18...

Page 36: ...4957D E F Microwave Analyzer Quick Start Guide 32 4957D H32 Soft backpack 4957D H33 Aluminum alloy carrying case 4957D H34 Waterproof case...

Reviews: