4957D/E/F Microwave Analyzer Quick Start Guide 

Chapter III Typical Applications 

4957D/E/F  offers  a  variety  of  measurement  modes.  This  chapter  will  provide  a  detailed 
introduction to commonly used test cases in each mode. 

3.1 Calibration in the VNA Mode 

In  order  to  ensure  the  accuracy  of  the  measurement  results,  the  instrument  must  be 
calibrated  before  testing  to  eliminate  the  systematic  error  of  the  analyzer.  4957D/E/F 
adopts mechanical calibration method with the process is as follows: 

1. Frequency response error correction 

You can eliminate the frequency responses of the following measurements for measured 
test component assembly: 

  Reflection measurement 

  Transmission measurement 

a)  Frequency response error correction of reflection measurement 

1)  In the VNA mode, select the desired type of measurement: 

  If  reflection  measurement  is  required  at  port  1  (positive,  S11),  keep  the  default 

setting of the instrument. 

  If  reflection  measurement  is  required 

at  port  2  (reverse,  S22),  press  [Measure]  → 

[S22]. 

2)  Set other parameters of the equipment measurement: Start Freq, Stop Freq, Power, 

Points or IF BW, etc.. 

3)  Press [Cal] to open the calibration menu. 

4)  Press [Cal Kit] to see if the calibration kit is consistent with the kit specified with the 

current  soft  key.  If  not,  press  the  corresponding  soft  key  to  select  the  correct 
calibration kit model. 

5)  Press  [OK],  return  to  the  menu  bar  of  [Cal],  select  the  frequency  response  error 

correction function, press [M Cal>] → [Freq Resp>] → [OPEN S11>], and the system 
will automatically prompt "Connect OPEN in P1, then press corresp. menu button to 
start!" then connect the open circuit device to port 1 selected as the test port. 

6)  After the display trace is  stable, to measure the standard piece, press [OPEN], the 

analyzer will display "calibration standard measurement …" during the measurement 
of  the  standard  piece,  and  after  completing  the  measurement  and  calculating  the 
error  coefficient,  it  will  display  the  prompt  "Press  [Done]  to  complete  the 
measurement!" And the software menu will be underlined to indicate that the project 
has been tested, such as [OPEN]. 

7)  Press  [Done]  to  complete  the  error  correction  of  reflection  measurement,  which 

allows for connection and measurement of the DUT.  

8)  Similarly, repeat step 5), select another frequency response error correction function, 

press  [M  Cal>]  →  [Freq  Resp.>]  →  [SHORT  S11>]  or  [OPEN  S22>]  or  [SHORT 
S22>], and connect an open circuit or a short circuit device at the corresponding test 
port according to different prompts to perform steps 6) and 7). 

 

The standard allows only one standard piece to be measured. If a 
wrong key is pressed when selecting a standard piece, return to 
[Cal] → [M Cal] → [Freq Resp.] and select the correct standard 
piece. Do not use the standard pieces of the circuit closer in 
reflection response correction. 

Note 

Summary of Contents for 4957D

Page 1: ...4957D E F Microwave Analyzer Quick Start Guide China Electronics Technology Instruments Co Ltd...

Page 2: ......

Page 3: ...duct can meet the indicator requirement in the manual upon delivery The calibration and measurement are completed by measuring bodies with national qualification with relevant data to be provided for...

Page 4: ...A Mode 9 3 2 DTF Measurement in the CAT Mode 15 3 3 Signal Measurement in the SA Mode 16 3 4 Measuring Signal Power in the Power Monitor PM Mode 18 3 5 Measuring the Electrical Length of Cables or Oth...

Page 5: ...the following chapters and sections Get Prepared This chapter introduces the preparation before operation startup panel introduction and battery replacement of 4957D E F After reading this chapter you...

Page 6: ...the instrument not mentioned in the manual Otherwise such consequences as reduced electromagnetic shielding property and internal component damage can occur affecting product reliability In this case...

Page 7: ...e The basic parameters of the enclosed battery are as follows Nominal voltage 10 8 V Nominal capacity 7 800 mAh Working time About 2 5 h Charging time About 6 h Rechargeable batteries should not be ex...

Page 8: ...s on the bottom of the touch screen are represented by XXX with XXX being the name of the key The menu keys on the right are represented by XXX with XXX being the menu name 2 3 Front Panel This sectio...

Page 9: ...ch can be selected as CAT Network Analyzer Spectrum Analyzer Power Monitor USB Power Meter and Voltage Voltmeter etc Cal It is used to select the type of device under calibration piece perform calibra...

Page 10: ...installed power connected b Battery installed and full power connected Flashing yellow a Battery installed and not full power connected Power on Normal on green a Battery not installed power connected...

Page 11: ...arked as 10 MHz In Out with 50 impedance BNC female adapter 4957D E F microwave analyzers can receive externally input 10 MHz sine signals as phase locked reference signals and the input signals shoul...

Page 12: ...ally switched from the speaker to the earphone 2 5 4 Instrument Symbols The instrument symbol yellow label shown in the figure indicates that the max power input from the test spectrum input port and...

Page 13: ...g soft key to select the correct calibration kit model 5 Press OK return to the menu bar of Cal select the frequency response error correction function press M Cal Freq Resp OPEN S11 and the system wi...

Page 14: ...pt Connect Thru cable between P1 and P2 then press corresp menu button to start Then connect N type cable between ports 1 and 2 All adapters or cables used in device measurement should be included i e...

Page 15: ...location where the DUT is to be connected 5 After the display trace is stable press THRU and the analyzer will display Cal std measuring during the measurement of the standard piece and after completi...

Page 16: ...etc 3 Open the calibration menu and press Cal 4 Press Cal Cal Kit to see if the calibration kit is consistent with the kit specified with the current soft key If not press the corresponding soft key t...

Page 17: ...the breaker and connect the load to the test port 10 After the display trace is stable to measure the standard piece press LOAD the analyzer will display calibration standard measurement during the me...

Page 18: ...e and connect the short circuit device to the test port 1 After the display trace is stable to measure the standard piece press P1 SHORT the analyzer will display calibration standard measurement duri...

Page 19: ...layed in the information zone Connect the DUT for measure Figure 3 5 Standard piece connection of full 2 ports error correction In actual calibration there is no sequence for the open circuit device s...

Page 20: ...w to display the submenu of Window and then select different window functions 5 Press Back Unit m ft to set the required distance unit system 6 Press Cable Specification and Load Cable to select the m...

Page 21: ...ency value 4 Setting sweep span Press Freq Span Note whether the span data is displayed in the information input zone to determine the currently activated parameters Reduce the span to 20MHz for insta...

Page 22: ...eed of connection to an external power probe can conveniently test the power of continuous wave signals with a certain bandwidth and the channel bandwidth can be set to a larger value so as to simulat...

Page 23: ...onvenient operation interface as the vector voltmeter The results of single port electrical length measurement are displayed in amplitude and phase formats while the results of all port are only displ...

Page 24: ...port error correction in the VVM mode 5 Selecting measurement mode The VVM mode of the analyzer allows for measurement of parameters of the DUT like amplitude phase VSWR and impedance etc 1 Press Mea...

Page 25: ...USB Power Meter mode and press OK to complete Figure 3 11 Test interface of USB Power Meter mode 2 Connecting power probes 1 As shown in the figure connect the small end of the USB cable to 8723X ser...

Page 26: ...required to set the frequency value of the measured CW Freq signal The USB power probe will store a set of correction factors for each frequency The frequency corresponding to the measured power signa...

Page 27: ...e losses for added attenuators or connecting cables It also allows for setting power offsets for gains of added amplifiers Positive values can compensate losses and negative values can compensate gain...

Page 28: ...F microwave analyzer provides two limit lines upper and lower When the amplitude of a signal in a certain frequency band exceeds the set upper limit or is less than the set lower limit the analyzer w...

Page 29: ...rmat or csv data text format The Print Screen function allows for saving the information displayed on the current screen into the memory of the instrument in an image format The storage medium for sta...

Page 30: ...environment for compliance with the requirements the test port connectors for any damages and the calibration piece for normal performance indicators If everything above is normal the instrument may b...

Page 31: ...g to the following steps 1 Write a paper document describing the instrument failure and put it in the package together with the analyzer 2 Pack it with the original packing materials so as to minimize...

Page 32: ...y 32 dB 40 dB 32 dB 40 dB 28 dB 35 dB Effective source matching 25 dB 37 dB 25 dB 37 dB 18 dB 30 dB Reflected tracking 0 10dB 0 14dB 0 10dB 0 25dB 0 10dB 0 25dB Transmission tracking 0 25dB 0 33dB 0 2...

Page 33: ...nput level 27 dBm Test port type N type female 3 5 mm male 2 4 mm male Power supply Rechargeable li ion battery or power adapter Power consumption 45 W excluding battery charging Working temperature r...

Page 34: ...a test software 4957D S03 Vector voltmeter software 4957D S04 USB power measurement software USB power probe to be equipped separately 4957D S05 Power meter software 4957D S06 Field strength measureme...

Page 35: ...female male calibration cable GORE 0K0CJ0CK024 0 4957D H14 87230 USB CW power probe 4957D H15 87231 USB CW power probe 4957D H16 87232 USB CW power probe 4957D H17 87233 USB CW power probe 4957D H18...

Page 36: ...4957D E F Microwave Analyzer Quick Start Guide 32 4957D H32 Soft backpack 4957D H33 Aluminum alloy carrying case 4957D H34 Waterproof case...

Reviews: