ENG
2016-11-21
Page 21 of 44
CertoClav Sterilizer GmbH
Georg Grinninger-Str. 37
A-4050 Traun / Austria
Technical data
Measures and weight
Material
Stainless steal
304L
Certificates
CE
Dimensions
Chamber Volume
18 Litre
Device height
60 cm
Chamber diameter
26 cm
Device width
44 cm
Chamber height
35 cm
Device depth
40 cm
Chamber usable height
30 cm
Device weight
15 kg
Operating data
Max. operating
temperature
140 °C
Power supply
230V, 50/60 Hz
Max. operating pressure
2,7 Bar (39 psi)
Power
1300 W
Test pressure
4,1 Bar (59 psi)
Heating duration
Amount of load
Heating from 25 °C to 121 °C
Little
15
– 20 Minutes
A lot
25
– 40 Minutes
Cooling duration
Amount of load
Cooling duration at faster
steam exhaust from 121 °C to
80°C
Cooling duration at slower steam
exhaust from 121 °C to 80°C
Little
20
– 35 minutes
40
– 60 minutes
A lot
40
– 60 minutes
80
– 120 minutes
Correlations table