![CenterVue COMPASS Operating Manual Download Page 49](http://html1.mh-extra.com/html/centervue/compass/compass_operating-manual_2581636049.webp)
CenterVue
COMPASS Operating Manual
Pag. 49 di 53
Manufacturers EMC Declaration to ISO 60601-1-2
The following tables provide specific information regarding compliance of COMPASS. COMPASS
is intended for use in the electromagnetic environment specified below. The customer or the user of
COMPASS should ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
COMPASS uses RF energy for its internal function.
Therefore, its RF emissions are very low and not likely
to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
COMPASS is suitable for use in all establishments,
including domestic and those directly connected to the
public low-voltage supply network that supplies
buildings used for domestic purposes, providing the
following warning is heeded:
Warning
: This equipment/system is intended for use
by healthcare professionals only. This
equipment/system may cause radio interference or
may disrupt the operation of nearby equipment. It
may be necessary to take mitigation measures, such as
re-orientating or re-locating COMPASS or shielding the
location.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions IEC
61000-3-3
Complies
Table 1: Electromagnetic Emissions