6-1 Accuracy Specification
Notes:
Inductance @ Ta =18 ~ 28
(De)
C
Measurement performed at the test socket.
Measurements performed after correct open and short
calibration.
DUT and test leads must be properly shielded to guard if
necessary.
Q value is the reciprocal of DF.
Accuracies based within 10% to 100% of full scale of range;
values outside of range should be used as reference only.
— means parallel or series measurement mode.
Frequency = 100 Hz/120 Hz
Frequency = 1kHz
O
•
•
•
•
•
•
Range
20.000mH
200.00mH
2000.0mH
20.000H
200.00H
2000.0H
20.000kH
Resolution
1uH
0.01mH
0.1mH
1mH
0.01H
0.1H
0.001kH
Lx Accuracy
1.5% 10d
1.4% 15d
1.5% 15d
1.6% 10d
1.3% 10d
2.0% 15d
2.5% 15d
±
±
±
±
±
±
±
DF Accuracy
1.5% 50d
1.4% 50d
1.5% 50d
1.6% 50d
1.3% 50d
2.0% 50d
2.5% 0d
±
±
±
±
±
±
±
Measurement
Mode
Series
Series
Series
---
Parallel
Parallel
Parallel
Range
2000.0uH
20.000mH
200.00mH
2000.0mH
20.000H
200.00H
2000.0H
Resolution
0.1uH
1uH
0.01mH
0.1mH
1mH
0.01H
0.1H
Lx Accuracy
1.3% 10d
1.2% 10d
1.2% 10d
1.5% 15d
1.5% 15d
2.0% 10d
2.5% 15d
±
±
±
±
±
±
±
DF Accuracy
1.3% 50d
1.2% 50d
1.2% 50d
1.5% 50d
1.5% 50d
2.0% 50d
2.5% 50d
±
±
±
±
±
±
±
Measurement
Mode
Series
Series
Series
---
Parallel
Parallel
Parallel
15
Dual Display LCR Meter Instruction Manual