Discover Bio
™
Manual Microwave Peptide Synthesizer
600423 • Revision 2 • August 2017
6
2.4. Reconnect the reaction vessel body to the top by pressing the body up into the top and
rotating.
3. Connect the thermowell.
3.1. Place a yellow ferrule and nut on the end of the thermowell.
3.2. Insert the thermowell into one of the side ports of the reaction vessel top so that the bottom
of the thermowell is just resting on the bottom of the vessel. Secure the thermowell with
the nut.
4. Connect the reaction vessel vent line.
4.1.
Cut a length of tubing (~48” (120 cm)). Using a yellow ferrule and nut, connect the tubing to
the other side port of the reaction vessel top, being careful to bend the sprayhead line out of
the way.
4.2.
Using a yellow ferrule and nut, connect the other end of the tubing to position 9 on the rear
of the Wash Station
5. Connect the RV Drain line.
CAUTION
Do not overtighten the fitting on the thermowell, as this can break the thermowell and damage
the fiber-optic probe.