System Requirements
Bench and/or Fume Hood Space
The Discover Bio should be positioned on the bench such that access to the electrical outlets for the system is not
restricted. The Discover Bio requires the following space for system components:
Discover Bio Microwave Reactor
15” (w) x 20” (d) x 9” (h) [38 cm (w) x 51 cm (d) x 23 cm (h)]
(Depth includes 3” (8 cm) clearance behind instrument for unimpeded airflow at rear fan ducts)
Discover Bio Wash Station
11” (w) x 11” (d) x 13” (h) [28 cm (w) x 28 cm (d) x 33 cm (h)]
Bottle Holder with Bottles
9.5” (w) x 9” (d) x 11” (h) [24 cm (w) x 22 cm (d) x 28 cm (h)]
Environmental Conditions
The Discover Bio is designed for indoor used only.
Temperature Range: 50 °F – 85 °F (10 °C – 29 °C)
Relative Humidity Range: 0 – 85%
Inert Gas Source
The Discover Bio requires an inert gas source (either high purity grade nitrogen or argon) capable of supplying 20
psi (20 L/min flow) within ten feet (10’)/three meters (3 m) of the rear of the instrument.
Electrical Requirements
The Discover Bio requires electrical power of 120 VAC (60 Hz, 1.7 A) (or 240 VAC [50 Hz, 1.7 A] where applicable).
Specific power requirements (120 VAC vs. 240 VAC) can be found on the nameplate affixed to the rear of the
Discover Bio Microwave Reactor and on the rear of the Discover Bio Wash Station.
Safety Information
Fume Ventilation
The Discover Bio operates as a semi-closed system, with minimal venting of any hazardous solvent fumes through
the vent line. The vent line must be vented into a proper chemical fume hood or exhaust line no longer than 10
feet (10’)/three meters (3 m) from the instrument.
In addition, adequate ventilation should be provided for preparation of reagents and solvents for use on the
system. All solvent bottles and the Waste Reservoir should be placed into proper secondary containers to minimize
the risk of exposure.
Waste Disposal
Waste produced by the Discover Bio can be hazardous. For detailed information on the safety requirements for the
chemicals used on the Discover Bio, refer to the appropriate SDS documents.
WARNING
Handle all waste under a fume hood, and wear suitable protective clothing. Dispose of all waste
in accordance with all applicable local, state, and federal heath and safety recommendations.