User Manual
Calibration and
Offset Instructions
Turn on the Slide Drying Hotplate.
Set the temperature of the Slide Drying
Hotplate to the desired temperature.
Let the instrument reach working
temperature and leave for 30 minutes
before carrying out a calibration check,
this ensures that the temperature
of the hotplate has equilibrated.
Check temperature with a calibrated
digital thermometer fitted with a surface
probe. The temperature should be
measured in the centre of the hotplate
(back to front and side to side)
, as shown
in the image.
Compare the temperature shown on
the Slide Drying Hotplate display with
the temperature shown on your external
digital probe.
If the two numbers are in close
agreement i.e. within ± 1°C
(as shown in
the image above)
, then the instrument
is within calibration. If the temperatures
are not in agreement i.e. a difference
greater than ± 1°C, then the instrument
will need to be re-calibrated.