CellPath JAY-0100-00A User Manual Download Page 6

www.cellpath.com   |   T: +44 (0)1686 611333

6

User Manual

Operating Instructions 

Œ

  

Place the hotplate on a smooth and level 

work surface.



  

Place the four microscope slide support 

rods in position according to user 

preference.

Ž

  

Connect the mains plug to the electrical 

supply and switch on 

(Ensure the power 

supply is properly earthed)

.



  

Turn on the hotplate.



  

Select the desired temperature.

A. Press button 

P then release it (

do not 

hold down button 

P

 for 5 seconds)

.

B. The display will show SP alternating 

with the current set temperature.
C. To change the set temperature 

press the 

UP key to increase the value 

or 

DOWN to decrease it. These keys 

increase or decrease the value one 

digit at a time, but if the button is 

pressed for more than one second 

the value increases/decreases rapidly, 

and after two seconds pressed, the 

speed increases even more to allow the 

desired values to be reached rapidly.
D. Exiting the Set mode is achieved by 

pressing the 

P key or automatically if 

no key is pressed for 15 seconds.  

After that time the display returns  

to the normal function mode.

‘

  

The heater indicator will illuminate to 

show heater activity.

’

  

The instrument will then warm up to the 

desired temperature, you will observe 

the temperature rise on the display.

“

  

The instrument is designed to warm  

up quickly 

(roughly 3 to 5 minutes)

.  

This means that when initially warming 

up, the instruments will slightly over 

shoot the set target temperature  

(by 2-5°C)

. The heater will then stop 

and allow the instrument to cool to the 

desired working temperature and then 

maintain the set working temperature 

at +/-1°C, by alternating heating and 

cooling.

Summary of Contents for JAY-0100-00A

Page 1: ...Slimline Slide Drying Hotplate User Manual www CellPath com T 44 0 1686 611333...

Page 2: ...86 611333 2 User Manual Contents 3 General Notes 3 Specimen Safety 4 User Safety 4 Power Lead and Connection to Electrical Supply 5 Specification 5 Location 6 Operating Instructions 7 Cleaning Instruc...

Page 3: ...should be replaced with an equivalent power supply unit Before using any cleaning or decontamination method please refer to the Maintenance and Cleaning section to ensure the proposed method will not...

Page 4: ...h will prolong the life of the instrument to ensure continued safety Do not touch any electrical contacts or open any closure plates RISK OF ELECTRIC SHOCK DO NOT Allow molten wax to accumulate on the...

Page 5: ...limline Slide Drying Hotplate features a digital temperature control over temperature cut out and miniature circuit breakers for safety Dimensions Width 183mm Depth 450mm Height 85 5mm Capacity 28 Sli...

Page 6: ...ssed for more than one second the value increases decreases rapidly and after two seconds pressed the speed increases even more to allow the desired values to be reached rapidly D Exiting the Set mode...

Page 7: ...l quantity of mild domestic detergent applied with a soft synthetic sponge The microscope slide support rods can be removed to facilitate easier cleaning SCOURING PADS OR DE SCALING AGENTS MUST NOT BE...

Page 8: ...ck temperature with a calibrated digital thermometer fitted with a surface probe The temperature should be measured in the centre of the hotplate back to front and side to side as shown in the image C...

Page 9: ...splay will now show a number this number is the current adjustment offset applied to the temperature display This was the adjustment offset which was used when the slide drying hotplate was manufactur...

Page 10: ...Thermocouple PTFE Insulated 1 Off Routine Inspection Recommendations CellPath Ltd recommend that a simple annual inspection be made for all CellTec laboratory equipment in order that any malfunction...

Page 11: ...1333 11 Slimline Slide Drying Hotplate Wiring Diagram SW1 MCB x2 E N L BODY HOB R38 ELEMENT 1 ELEMENT 2 NEON 1 3 SSR 5 6 4 2 110V N AC AC DC DC N N 230V CUT OUT 120 C 1 2 3 4 5 6 7 8 11 12 OUT 2 OUT 1...

Page 12: ...e 2 Power is supplied to the instrument but the hot plate does not heat Temperature does not rise on the controller and the orange heater light does not operate A Thermal cut out has triggered B Tempe...

Page 13: ...instrument may need to be re calibrated Follow the calibration instructions A Check correct probe is being used for measurement and that the probe is calibrated B Measure temperature at the position w...

Page 14: ...to or inscribed upon the goods have been removed covered up or defaced in any way Any goods or parts thereof which may require repair or replacement shall be repaired or replaced at the election of Ce...

Page 15: ...ection and repair of the equipment will be issued prior to any work being carried out On return of the repaired equipment to the customer it is the customer s responsibility to ensure that the loan eq...

Page 16: ...www cellpath com T 44 0 1686 611333 16 User Manual Notes...

Page 17: ...17 www cellpath com T 44 0 1686 611333 17 Slimline Slide Drying Hotplate Notes...

Page 18: ...www cellpath com T 44 0 1686 611333 18 User Manual Notes...

Page 19: ...w Voltage Directive 2014 35 EU EMC Directive 2014 30 EU We confirm the declaration Nickel Electro Ltd Oldmixon Crescent Weston Super Mare North Somerset BS24 9BL United Kingdom Conforms with the requi...

Page 20: ...CellPath Ltd 80 Mochdre Enterprise Park Newtown Powys SY16 4LE UK T 44 0 1686 611333 E sales CellPath com CellPath com 00 659 372 V3...

Reviews: