User Manual
Operating Instructions
Place the hotplate on a smooth and level
work surface.
Place the four microscope slide support
rods in position according to user
preference.
Connect the mains plug to the electrical
supply and switch on
(Ensure the power
supply is properly earthed)
.
Turn on the hotplate.
Select the desired temperature.
A. Press button
P then release it (
do not
hold down button
P
for 5 seconds)
.
B. The display will show SP alternating
with the current set temperature.
C. To change the set temperature
press the
UP key to increase the value
or
DOWN to decrease it. These keys
increase or decrease the value one
digit at a time, but if the button is
pressed for more than one second
the value increases/decreases rapidly,
and after two seconds pressed, the
speed increases even more to allow the
desired values to be reached rapidly.
D. Exiting the Set mode is achieved by
pressing the
P key or automatically if
no key is pressed for 15 seconds.
After that time the display returns
to the normal function mode.
The heater indicator will illuminate to
show heater activity.
The instrument will then warm up to the
desired temperature, you will observe
the temperature rise on the display.
The instrument is designed to warm
up quickly
(roughly 3 to 5 minutes)
.
This means that when initially warming
up, the instruments will slightly over
shoot the set target temperature
(by 2-5°C)
. The heater will then stop
and allow the instrument to cool to the
desired working temperature and then
maintain the set working temperature
at +/-1°C, by alternating heating and
cooling.
Summary of Contents for JAY-0100-00A
Page 1: ...Slimline Slide Drying Hotplate User Manual www CellPath com T 44 0 1686 611333...
Page 16: ...www cellpath com T 44 0 1686 611333 16 User Manual Notes...
Page 17: ...17 www cellpath com T 44 0 1686 611333 17 Slimline Slide Drying Hotplate Notes...
Page 18: ...www cellpath com T 44 0 1686 611333 18 User Manual Notes...