48
OPERATOR'S MANUAL
EN
Afterwards, let the patient in and position him/her as
described in paragraph 6.2 "Positioning the patient for
3D exams".
By clicking on the arrows in the right-hand box it is
possible to improve the column positioning, moving it
upwards or downwards.
Once the patient has been correctly positioned, click
on the NEXT button.
7) Starting the scan
Choose between the PATIENT SCAN or DENTURE
SCAN option to pass the control to the NNT software.
8) Scanning mode selection
In the following screen, select the Eco Scan or the
Regular Scan option in order to start the CBCT
acquisition process.
Follow the instructions for the acquisition operations
described in the NNT manual to complete image
acquisition.
Compared to the Regular Scan mode, the Eco Scan option has a lower dose and a smaller number of
acquired images for all modes, alongside a reduced scanning time for the Hi-Res modes and for the fields of
view with a diameter of 8 cm and 10 cm in standard mode.
Thanks to these characteristics, the Eco Scan option is especially recommended in place of the Regular
Scan mode for pediatric patients.