13. INSPECTION AND MAINTENANCE
13.1. USER INSPECTION
These instructions describe the maintenance procedures for the extraoral X-ray system.
These instructions apply to all versions of said equipment, as well as all the accessories that may have been provided, therefore the description of some parts may not correspond to your equipment.
Inspection and preventive maintenance must be performed at scheduled intervals to protect the health and safety of patients, users and other persons in accordance with national regulations regarding the use and maintenance of dental x-ray
units that are in force in the country where the device is installed.
In order to ensure the operational safety and functional reliability of your product, the system owner should check the equipment at regular intervals (at least once a year) or commission an authorised technician to do so.
If one or more checks to be performed are not satisfactory, please contact your dealer for technical support.
Answer questions with yes (√) or not (−)
Inspection DATE
Step
Description
Reference in the manual
__/__/20__
__/__/20__
__/__/20__
__/__/20__
__/__/20__
1
Check that all labels located
-
on the detector/s
-
on 3D panel (if any)
-
at the base of the column,
-
on the X-ray tube housing,
are intact, well attached and legible.
Section Position of identification labels
2
Check there are no external damages to the equipment, which may reduce protection against radiation.
Section Description of the operations
3
Check the remote emission button cable is not broken or scratched.
Section X-ray emission remote control
4
Check the power switch, verifying that it is working properly and that the touch panel display switches on
when the switch is in the ON position.
Section Starting the system
5
Verify that exposure is immediately interrupted when x-ray button is released
Section X-ray emission remote control
6
Check the touch screen console functionality: functions must respond to interrogation.
Section Touch screen display
7
Check proper functioning of X-ray exposure LED and exposure buzzer
Section X-ray emission remote control
8
Check proper functioning of X-ray centring laser traces
Section Laser positioning system
9
Make sure that the detector is smoothly inserted inside the guides on Pan and CEPH sides - only for
units with Cephalometric arm and removable detector.
Section Sensor positioning
10
Perform a movement test running the Dummy panoramic procedure, making sure the movement on the
3 axes X, Y, R (rotation and translation simultaneously) is smooth and without noise. See figure 1, next
page.
To perform a dummy cycle, namely the simulation of an examination without the emission of
X-rays, select any type of panoramic X-ray examination and reset the unit. Once the reset
position has been reached, press the “Reset” button and keep it pressed during the entire
procedure, otherwise the movement will stop.
11
If a Cephalometric arm is installed, perform a movement test for the cephalometric detector using the
Dummy Ceph test procedure, and making sure that the movement on the H axis (translation) is smooth
and without noise. See figure 1, next page.
To perform a dummy cycle, namely the simulation of an examination without the emission of
X-rays, select any type of cephalometric examination and reset the unit. Once the reset
position has been reached, press the “Reset” button and keep it pressed during the entire
procedure, otherwise the movement will stop.
12
Perform column movements, up and down on the Z axis and check proper functioning. See figure 2,
next page.
Section Control panel (console onboard the machine)
13
Check the emergency stop button functionality. Emergency stop button is used to stop X-ray device
operation, it is located under the patient’s support arm, near the telescopic column (
Emergency stop
button)
User Manual, section Emergency stop button
14
Check the x-ray generator functionality performing a complete trial exposure. Select any panoramic
exam and reset the unit. Hold down the emission button throughout the entire exam procedure. Absence
of error messages assures proper generator functionality.
Section Performing a 2D X-Ray examination
If Quality Phantoms are not available at installation site, then contact your dealer tech support to have the Quality Assurance procedure performed at your premises
15
Perform an exam on the 3D Quality Phantom and evaluate the quality of the outcome volumetric study.
Acquisition Operations with Annex
Operator Name
Signature
The undersigned confirms that the equipment was checked for the above criteria and that, in case of any malfunction, an authorised technician of the local dealer was informed.
All inspection and maintenance work performed by the system owner and/or service engineer must be recorded in this document and kept near the unit!