Code
Message
Solution
0.13
2D Detector failed to turn ON
Check the removable sensor
’s
contacts. Unplug the sensor and
plug it back.
0.14
X-Ray Start-Command aborted
1) Check
Converter
board
diagnostic LEDs (>15s).
2) Verify CAN connector (KCB1) on
converter board.
0.16
Relocate 2D Detector to PAN
position
Relocate 2D Detector to PAN.
0.17
Relocate 2D Detector to CEPH
position
Relocate 2D Detector to CEPH
position
0.18
Check correct position of Craniostat
Check correct position of Craniostat
0.19
2D Detector is not recognized (run
"Sensor registration")
Please serialize the sensor (from
unit calibration software on PC)
0.20
2D Detector not calibrated
Calibrate the sensor (from unit
calibration software on PC)
0.21
Serial Number Mismatch: not the
same 2D Detector used for
calibration
Please serialize the sensor (from
unit calibration software on PC)
0.22
Detected Open Door (interlock is
active)
Check interlock switch (door not
closed?)
0.23
Image
not
downloaded
from
equipment, transfer to PC or USB
before proceeding
New exposure requested while
previous has not been downloaded
yet
0.24
Please wait for confirmation before
press
The xray button was pressed before
starting the cbct exam procedure.
For CBCT exams, launch “Patient
scan” from within iRYS and wait for
an explicit request before pressing
the xray button.
0.25
CEPH enabled
Only at first startup after CEPH arm
installation, call tech support if
persistent.
0.26
CMOS Time and Date Not Set.
CMOS battery low?
Set time and date from PC/Check if
clock battery on mainboard is dead
0.27
Device not configured
Set configuration from PC
0.28
Device not calibrated
Perform unit calibration
0.29
Please, let the patient release the
bite, then press reset to move the
chinrest to the new position
Chinrest is going to perform a long
movement, so the patient must
release the bite and let the chinrest
free to move
0.30
Buttom pressed on startup
Detected buttons pressed during
startup. Verify that all the keyboard
buttons are working properly.
0.31
USB Key not found
0.32
USB error on writing
USB key not compatible / Corrupted
/ Full
0.33
Unable to move detector's carousel
automatically. Please, move it
manually