·   The electrodes begin to lose their stickiness and do not stick properly to the 

skin. Moisten the adhesive surface with a few drops of water before placing 
on the skin.

·   The electrodes are worn out and need to be replaced.

·   Slightly change the position of the electrodes.

The stimulation feels weak or not at all

·   Check if the batteries need to be replaced, see section 

Batteries

.

·   Electrodes are too old and need replacement. 

The broken circuit symbol is shown on the display 

 

The broken circuit symbol indicates that the resistance is too high, or that a 
cable is broken.

·   A too high resistance can be caused by a bad connection between the elec-

trodes and your skin, or that the electrodes need to be replaced.

·   A cable breakage can be checked by pressing the cable’s pins against one 

another while increasing the amplitude for the corresponding channel to 
11 mA. If the amplitude now drops to 0.0 mA and starts flashing, the cable 
needs to be replaced.

   Note!

 Never increase the amplitude above 20 mA when you check for cable 

breaks, since this can damage the stimulator.

TEChNICAL DATA

CEFAR MYO X2 and CEFAR ACTIV X2 

are electrical stimulators with two chan-

nels intended for muscle training and pain relief. They can deliver 100 mA of 
constant current through a resistance load of 1000 ohms. With a change in 
load from 100 to 1000 ohms, the stimulation current changes less than 10 % 
from the set value. The stimulator operates on either two non-rechargeable 1.5 
V AA batteries or two rechargeable 1.2 V AA batteries, recharged in a separate 
charger.

Information Related to Electromagnetic Compatibility (EMC)
CEFAR MYO X2 and CEFAR ACTIV X2 

are designed to be used in typical 

domestic or commercial environments and are approved according to the EMC 
safety standard of EN 60601-1-2. 

CEFAR MYO X2 and CEFAR ACTIV X2

 emit very low levels in the radio fre-

quency (RF) interval. Therefore they are not likely to cause any interference in 
nearby electronic equipment (radios, computers, telephones etc.).

CEFAR MYO X2 and CEFAR ACTIV X2

 are designed to withstand foreseeable 

disturbances originating from electrostatic discharges, mains supply magnetic 
fields and radio frequency transmitters (such as mobile telephones).

2526_CEFAR_MYO.ACTIV_X2__ENG_3-113   13

2006-07-07   11:10:00

Summary of Contents for activ X2

Page 1: ...i 27 Manual de instrucciones 39 Istruzioni per l uso 51 Manual de instru es 63 Gebruikershandleiding 75 87 Instrukcja obs ugi 99 111 Haszn lati tmutat 123 bruksanvisning 135 istimal sureti 147 bruksan...

Page 2: ...ual Amplitude Setting 8 Program Guide 9 CEFAR EASY TRAINING Sets and Repetitions Reps 9 Active Rest 9 Levels 9 Program Description 9 CEFAR MYO X2 9 CEFAR ACTIV X2 11 Key Lock 12 Pausing the Program 12...

Page 3: ...regnant The reason is that theoretically the current could affect the foetus s heart although there are no reports of it being harmful If skin irritation should occur treatment should be temporarily d...

Page 4: ...cted amplitude level in AUTO test mode 3 AUTO P AUTO S Decreases or increases the stimulation amplitude intensity of the stimulation for each channel Note Always increase the amplitude cautiously AUTO...

Page 5: ...Number Intensity level See Program Description for further information 8 remaining Time Remaining program time in minutes and seconds Time is flashing during timer setting also counts down the time f...

Page 6: ...2 Attach the electrodes to your body Follow the recommendation in Electrode Placement Guide at the end of this user manual 3 Connect the cable to one of the stimulator s two channels 4 Press AUTO P A...

Page 7: ...o increase or decrease stimulation amplitude at any time until the stimulation feels pleasant presumed that the key lock is inactivated If doing so the Automatic Amplitude Setting is lost and the stim...

Page 8: ...decrease muscle soreness afterwards Note Do not forget to set the amplitude for the Active Rest in manual mode In auto mode the amplitude adjusts automatically in Active Rest Levels Some of the progra...

Page 9: ...Improves circulation and elimination of lactic acid after a work out Note The TENS program does not always treat the cause of the pain If the pain persists seek medical advice Active Recovery P21 Inc...

Page 10: ...aximal training ses sions P13 lvl 3 Warm Up P1 Prepares the muscle for exercise Recovery P14 Improves circulation and elimination of lactic acid after a work out Drainage P15 The comfortable vibration...

Page 11: ...e AA 1 5V Make sure the polarisation placement and is correct 4 Replace the battery cover 5 Dispose of the exhausted batteries according to local regulations Care Instructions Taking care of and clean...

Page 12: ...nce this can damage the stimulator Technical Data CEFAR MYO X2 and CEFAR ACTIV X2 are electrical stimulators with two chan nels intended for muscle training and pain relief They can deliver 100 mA of...

Page 13: ...chargeable or 2 x 1 2 V AA rechargeable batteries Current consumption for one channel 80 Hz 30 mA 150 mA I r m s max channel 11 mA Size 120 x 50 x 30 mm Weight ca 180 g Key to the Stimulator Symbols R...

Page 14: ...195 2526_CEFAR_MYO ACTIV_X2_EP_195 2195 195 2006 07 07 11 49 39...

Page 15: ...196 2526_CEFAR_MYO ACTIV_X2_EP_195 2196 196 2006 07 07 11 49 40...

Page 16: ...197 2526_CEFAR_MYO ACTIV_X2_EP_195 2197 197 2006 07 07 11 49 41...

Page 17: ...198 2526_CEFAR_MYO ACTIV_X2_EP_195 2198 198 2006 07 07 11 49 42...

Reviews: