background image

page 7

TORQUE MEASUREMENT

page 6

TORQUE MEASUREMENT

OUTPUT

USB Virtual COM Port

When the tester is connected to a computer’s USB port and turned
on,  Windows XP or Vista will recognize the tester and ask to con-
nect to the Internet. Allow Windows Update to connect to the
Internet and follow the instructions on your screen to download
and install two drivers for the USB virtual com port. For Windows
98, ME and MAC go to www.ftdichip.com/Drivers/VCP.htm.

The PC and tester will communicate through the USB port the same
as a com port.  To verify the COM port number open the Device
Manager.

Output is available in PD, PP and C modes.  In PD and PP modes peak
data is output when ZERO is pressed or activated by the AUTO ZERO
function. In C mode, the gauge outputs data continuously 180
data/second.

 

 
 

 

 
 
 
 

 

 

Office: Jl. Radin Inten II No. 62 Duren Sawit, Jakarta 13440 - Indonesia

Workshop: Jl. Pahlawan Revolusi No. 22B, Jakarta 13430 - Indonesia

Phone: 021-8690 6777 (Hunting)

Mobile: +62 816 1740 8925

Fax: 021-8690 6771

Summary of Contents for DI-1M

Page 1: ...ce department for a return authorization number and return the defective product to us with freight prepaid The foregoing warranty constitutes the SOLE AND EXCLUSIVE WARRANTY and we hereby disclaim all other warranties express statutory or implied applicable to the products and or software including but not limited to all implied warranties of merchantability fitness non infringement results accur...

Page 2: ...battery be sure to use the provided AC adapter charger exclusively l When measuring torque the unit will have a tendency to turn In order to avoid spinning and possible accident it is very impor tant to securely mount the unit to a stable work bench l Do not use lacquer thinner or any solvent to clean the unit l Do not disassemble or modify the unit Model lbf in kgf cm N m DI 1M 50 3 0 440 0 3 0 5...

Page 3: ...ecrease Auto Zero Reset Switch 1 2 3 4 5a 5b 6 7 8 9 10 12 11 1 2 3 4 5a 5 5b 6 7 8 9 10 11 12 DI 1M IP SPECIFICATIONS DI 1M IP50 DI 1M IP200 DI 1M IP500 Measuring Range 0 30 50 N m 3 0 200 N m 3 0 500 N m 3 0 500 kgf cm 30 2000 kgf cm 30 5000 kgf cm 3 0 430 lbf in 30 1800 lbf in 30 4500 lbf in Peak Down 0 70N m 7 0N m Operating Range 7 0kgf cm lbf in 70kgf cm lbf in Accuracy 0 5 F S Display 4 dig...

Page 4: ...nge even if a higher value is subsequently measured peak data output 4 Press ON OFF if the unit does not display zero select TR mode and use the Zero Adjustment Pot to make sure the display reads zero 5 Select Peak Mode PP or Peak Down mode PD 6 The blow counter counts the number of pulses for a specified time up to 9 9 seconds The counter works in PP mode only Set the time duration by using a poi...

Page 5: ... manually for each test the Automatic Zero Reset can be programmed by using a paperclip to push the Auto Zero Reset Switch As the Auto Zero Reset Switch is pressed use a pointed tool like a paperclip end the Automatic Zero Reset durations cycle 0 0C Þ 0 5C Þ 1 0C Þ 1 5C Þ 2 0C Þ 2 5C Þ 3 0C Select the desired duration For example 0 5C means 0 5 second and 0 0C means MANUAL RESET If a selection is ...

Page 6: ...indows 98 ME and MAC go to www ftdichip com Drivers VCP htm The PC and tester will communicate through the USB port the same as a com port To verify the COM port number open the Device Manager Output is available in PD PP and C modes In PD and PP modes peak data is output when ZERO is pressed or activated by the AUTO ZERO function In C mode the gauge outputs data continuously 180 data second Offic...

Reviews: