TORQUE MEASUREMENT

page 8

TORQUE MEASUREMENT

Auto Power Off

To maximize the life of the battery, power will automatically shut
off after 10 minutes of non-use.  

System Reset Button 

When battery has been depleted completely and a recharge has
been executed, the system may not yet work.  In this case, press the
System Reset Button 

.

Low Battery Indicator

“–L– –” will appear when tester needs to be recharged.  Turn off
power.

IMPORTANT!

Use the provided IMADA/CEDAR AC adapter/charger

exclusively and plug into the correct AC output.  It takes 8 hours to
charge fully.  Do not recharge for more than 12 hours.  When it is
fully charged, make sure the AC adapter/charger is disconnected to
avoid overcharging.

In order to avoid high heat, explosion or toxic fumes, please note
the following precautions:

Be sure to use the provided AC charger exclusively.
Do not plug the charger into unspecified higher voltage.
Do not recharge until LOBAT icon appears.

8

Auto Zero Reset Switch 

for Peak or Peak Down mode

Rather than resetting manually for each test, the Automatic Zero
Reset can be programmed by using a paperclip to push the Auto
Zero Reset Switch

.

As the Auto Zero Reset Switch 

is pressed (use a pointed tool,

like a paperclip end), the Automatic Zero Reset durations cycle.
(0.0C 

Þ 

0.5C 

Þ 

1.0C 

Þ 

1.5C 

Þ 

2.0C 

Þ 

2.5C 

Þ 

3.0C) Select the

desired duration. For example, 0.5C means 0.5 second and 0.0C
means MANUAL RESET.  If a selection is not made within 3 seconds,
the cycle begins again.

12

12

Virtual RS-232C Serial Output

Output is available in PD, PP and C modes.  In PD and PP modes peak
data is output when ZERO is pressed or activated by the AUTO ZERO
function. 

Comport signal

8 data, 2 stop, no parity. Baud Rate: 19,200 bps.

Output Connector

USB A/B  cable

Peak Data Output Format

[CAN] _ _ _ [SO] [value] _ [SI] [unit] [CR]

(_ _ _ is memory location)   

[CAN]:

ASCII control code 24

_:

Space (code 32)

[SO]:

ASCII control code 14

[value]:

Output data with sign and decimal point.  Plus sign
represents for CW torque and minus sign for CCW.
[Value] always occupy six locations and empty
locations will be filled with spaces.

[SI]:

ASCII control code 15

[unit]: 

N*m _ _ _=N.m
kgf*cm = kg.cm
lb*in _ = lb.in

[CR]:

ASCII control code 13 (Carriage Return)

Continuous Output Data Format

[CAN] [value] [CR]

Summary of Contents for DI-1M

Page 1: ...ce department for a return authorization number and return the defective product to us with freight prepaid The foregoing warranty constitutes the SOLE AND EXCLUSIVE WARRANTY and we hereby disclaim all other warranties express statutory or implied applicable to the products and or software including but not limited to all implied warranties of merchantability fitness non infringement results accur...

Page 2: ...battery be sure to use the provided AC adapter charger exclusively l When measuring torque the unit will have a tendency to turn In order to avoid spinning and possible accident it is very impor tant to securely mount the unit to a stable work bench l Do not use lacquer thinner or any solvent to clean the unit l Do not disassemble or modify the unit Model lbf in kgf cm N m DI 1M 50 3 0 440 0 3 0 5...

Page 3: ...ecrease Auto Zero Reset Switch 1 2 3 4 5a 5b 6 7 8 9 10 12 11 1 2 3 4 5a 5 5b 6 7 8 9 10 11 12 DI 1M IP SPECIFICATIONS DI 1M IP50 DI 1M IP200 DI 1M IP500 Measuring Range 0 30 50 N m 3 0 200 N m 3 0 500 N m 3 0 500 kgf cm 30 2000 kgf cm 30 5000 kgf cm 3 0 430 lbf in 30 1800 lbf in 30 4500 lbf in Peak Down 0 70N m 7 0N m Operating Range 7 0kgf cm lbf in 70kgf cm lbf in Accuracy 0 5 F S Display 4 dig...

Page 4: ...nge even if a higher value is subsequently measured peak data output 4 Press ON OFF if the unit does not display zero select TR mode and use the Zero Adjustment Pot to make sure the display reads zero 5 Select Peak Mode PP or Peak Down mode PD 6 The blow counter counts the number of pulses for a specified time up to 9 9 seconds The counter works in PP mode only Set the time duration by using a poi...

Page 5: ... manually for each test the Automatic Zero Reset can be programmed by using a paperclip to push the Auto Zero Reset Switch As the Auto Zero Reset Switch is pressed use a pointed tool like a paperclip end the Automatic Zero Reset durations cycle 0 0C Þ 0 5C Þ 1 0C Þ 1 5C Þ 2 0C Þ 2 5C Þ 3 0C Select the desired duration For example 0 5C means 0 5 second and 0 0C means MANUAL RESET If a selection is ...

Page 6: ...indows 98 ME and MAC go to www ftdichip com Drivers VCP htm The PC and tester will communicate through the USB port the same as a com port To verify the COM port number open the Device Manager Output is available in PD PP and C modes In PD and PP modes peak data is output when ZERO is pressed or activated by the AUTO ZERO function In C mode the gauge outputs data continuously 180 data second Offic...

Reviews: