TORQUE MEASUREMENT

page 4

TORQUE MEASUREMENT

GENERAL OPERATION

IMPORTANT!

Always make sure that the sensor is mounted securely

to a stable workbench to avoid spinning and possible accident.
1. Connect the torque sensor to the tester display unit with cable

provided.  When connecting the cable to the torque sensor,
rotate the round connector to find the matching groove, then
push in until it makes positive connection (When disconnecting
the cable, hold the silver metal part of the round connector and
pull out). 

2. Select measuring unit (lbf-in/lbf-ft, kgf-cm/ kgf-m, or N-cm/N-m).

If desired, affix unit designator to prevent it from being acciden-
tally changed.

3. Select among the following

Measurement Modes (For normal test-
ing, PP mode is recommended).

PP (Peak Mode):

Capture peak torque

PEAK Torque reading will not change
until a higher value is measured (peak
data output).

C (Real Time Mode):

Display torque transient (no output).

PD (Peak Down Mode):

Capture first peak value and will not

change even if a higher value is subsequently measured (peak data
output).

4. Press ON/OFF, if the unit does not display zero, select TR mode

and use the Zero Adjustment Pot. 

to make sure the display

reads zero. 

5. Select Peak Mode (PP) or Peak Down mode (PD).
6. The blow counter counts the number of pulses

for a specified time, up to 9.9 seconds. The counter
works in PP mode only. Set the time duration by
using a pointed tool like a paperclip to push the
switches 

, right to increase, left to decrease.

7.   Insert wrench drive securely into the sensor, hold

perpendicularly (not at an angle), and turn on
the tool to measure.

8. After taking measurement, press CLEAR 

to

zero the display for the next test.

11

3

4

SW-1SV-USB is a highly customized
Excel spreadsheet specifically
designed for capturing and analyz-
ing real-time data from Cedar
torque instruments with a USB
port. SW-1SV-USB provides a run-
ning log of all measurements along
with a graph and statistical infor-
mation including min, max, mean,
median, standard deviation, average
deviation and variance.

Requirements

1. PC computer with Windows XP or later and Excel 2000 or later.
2. CB-208 or CB-209 cable

Optional Data Acquisition Software SW-1SV-USB

Summary of Contents for DI-1M

Page 1: ...ce department for a return authorization number and return the defective product to us with freight prepaid The foregoing warranty constitutes the SOLE AND EXCLUSIVE WARRANTY and we hereby disclaim all other warranties express statutory or implied applicable to the products and or software including but not limited to all implied warranties of merchantability fitness non infringement results accur...

Page 2: ...battery be sure to use the provided AC adapter charger exclusively l When measuring torque the unit will have a tendency to turn In order to avoid spinning and possible accident it is very impor tant to securely mount the unit to a stable work bench l Do not use lacquer thinner or any solvent to clean the unit l Do not disassemble or modify the unit Model lbf in kgf cm N m DI 1M 50 3 0 440 0 3 0 5...

Page 3: ...ecrease Auto Zero Reset Switch 1 2 3 4 5a 5b 6 7 8 9 10 12 11 1 2 3 4 5a 5 5b 6 7 8 9 10 11 12 DI 1M IP SPECIFICATIONS DI 1M IP50 DI 1M IP200 DI 1M IP500 Measuring Range 0 30 50 N m 3 0 200 N m 3 0 500 N m 3 0 500 kgf cm 30 2000 kgf cm 30 5000 kgf cm 3 0 430 lbf in 30 1800 lbf in 30 4500 lbf in Peak Down 0 70N m 7 0N m Operating Range 7 0kgf cm lbf in 70kgf cm lbf in Accuracy 0 5 F S Display 4 dig...

Page 4: ...nge even if a higher value is subsequently measured peak data output 4 Press ON OFF if the unit does not display zero select TR mode and use the Zero Adjustment Pot to make sure the display reads zero 5 Select Peak Mode PP or Peak Down mode PD 6 The blow counter counts the number of pulses for a specified time up to 9 9 seconds The counter works in PP mode only Set the time duration by using a poi...

Page 5: ... manually for each test the Automatic Zero Reset can be programmed by using a paperclip to push the Auto Zero Reset Switch As the Auto Zero Reset Switch is pressed use a pointed tool like a paperclip end the Automatic Zero Reset durations cycle 0 0C Þ 0 5C Þ 1 0C Þ 1 5C Þ 2 0C Þ 2 5C Þ 3 0C Select the desired duration For example 0 5C means 0 5 second and 0 0C means MANUAL RESET If a selection is ...

Page 6: ...indows 98 ME and MAC go to www ftdichip com Drivers VCP htm The PC and tester will communicate through the USB port the same as a com port To verify the COM port number open the Device Manager Output is available in PD PP and C modes In PD and PP modes peak data is output when ZERO is pressed or activated by the AUTO ZERO function In C mode the gauge outputs data continuously 180 data second Offic...

Reviews: