5 Error messages 

Certain problems could be highlighted during the auto-test and are shown by an 
acoustic alarm and a code formed by LED’s associated to the reading cells. These 
signals do not allow the analytical operation and the technical service assistance 
department must be advised in order to check the problem. 
These signals are coded with a brief description about their significance. The RRV 
indication, for example, means that the LED associated to the reading cell is green 
while the other two are red. 
 

Code 

Description 

VRV 

Corrupted EEPROM analysis 

RRV 

Modification to the version of the data trace 

RVR 

Memory error of the personalisation analysis 

RVV 

Hardware error on the display 

VRR 

Non-identifiable error of another type 

 

NOTE! 

A specific case can occur when a particular sequence RRR occurs during 

the normal use of the instrument and not only during the start up. This 
sequence indicates overheating of the instrument and therefore a 
temperature of more than 41°C / 105.8°F. In this case it is necessary to 
switch off the instrument, let it cool down and then switch it on again. If the 

problem persist, <please contact the technical assistance department for 
repairing. 

 
In the following are summarized the instructions to be followed in case specific 
problems occur. For any other problem it is possible to contact the cdR technical 
assistance department. 
 

PROBLEM CHECK 
 

Problem 

Check 

The instrument does not 
switch on – the display 
does not show any 
information

 

After having checked the connection of the power plug, 
check if that there is a power supply at the input entry of the 
adapter (230 Vac) and also at the instrument (12Vdc). If 
necessary, substitute the external adapter with a similar one 
that has the same characteristics and check the polarity.

 

The instrument does not 
print

 

Check if the roll of thermal paper is present and correctly 
inserted.

 

It is not possible to enter 
into the analysis key 3 and 
into the absorption key 5

 

The instrument has not reached the operating temperature. 
20 minutes are necessary for the instrument to reach the 
temperature from switching on to being in an operational 
mode.

 

The readings are not 
aligned to historical data

 

Attention! The calibration procedure of the instrument is very 
simple but must be carried out correctly. Ensure that during 
the calibration the reading cells are not occupied by cuvettes 

Summary of Contents for FOODLABFAT

Page 1: ...Operator s manual October 2012 1 Instrument for the Quality Control of Oils and Fats In Vitro Analysis Made in Italy CDR s r l Firenze Italy Development Manufacturing Sales Technical Service...

Page 2: ...arch 2011 J Schenne 2nd edition R3 October 2012 J Schenne 3th edition USA Document made by Ing Buero fuer Bio Prozessanalytik J Schenne Ohlstuecken 9 21033 Hamburg Germany info foodlab arcor de In cas...

Page 3: ...and 6 1 3 2 Safety of the product 7 2 System description 9 2 1 The FOODLABfat analyser 9 2 2 Instrument nameplate 9 2 3 Operating parts 10 2 4 1 Key folio layout and display 10 2 3 2 Reading cells 10...

Page 4: ...f the results 16 4 6 Standardisation of tests 17 4 7 Menu description 17 4 7 1 Edit dialogue 17 4 7 2 Setup dialogue 18 4 7 3 Auto start dialogue 19 4 7 4 Printer dialogue 19 4 7 5 Info dialogue 20 5...

Page 5: ...way and using the correct methods as agreed by the manufacturer If the operator finds a disagreement between anything described in this manual and the instrument he must advise the manufacturer immed...

Page 6: ...rixes different from those indicated in the methodologies The instrument is specifically calibrated during construction to test the biological matrixes indicated in the methodologies The instrument ca...

Page 7: ...2 Safety of the product 1 3 2 1 European Directives consulted during the design stage The instrument has been designed and constructed following the health and safety requirements of the directive Tab...

Page 8: ...requisites of the norms EN 50081 1 Electromagnetic compatibility emission residential environment light industry EN 55022 A Radio disturbance measures on equipment for computer technology EN 50082 2...

Page 9: ...Reading cells Figure 2 1 Modules 2 2 Instrument nameplate The instrument has a nameplate that displays the CE mark and identification data Figure 2 2 Nameplate Important Never remove the nameplate fr...

Page 10: ...Up Move the cursor upwards on the display stop number of samples Down Move the cursor downwards on the display interruption of a running test Power ON OFF the instrument Numerical keys 0 9 Enter Confi...

Page 11: ...ep it clean don t touch it with fingers Figure 2 5 CDR test cuvette 2 3 4 CDR Reagent kit Test kit contains 10 pre filled cuvettes in an aluminium bag For example Test name Acidity Code 300125 Lot rea...

Page 12: ...in its carton must be stored in a closed environment within the temperature range 20 C to 70 C 4 F to 158 F and a relative non condensed humidity of 20 to 90 The storage place must not be exposed to c...

Page 13: ...ed as follows A Take the FOODLABfat out of the carton and put it on a plane surface in such a position that it is not exposed by mechanical vibrations or extreme temperature fluctuations or excessive...

Page 14: ...peration procedure The FOODLABfat is an instrument equipped with dedicated software for the deter mination of parameters on different oils and fat extracts and alimentary matrixes of various natures T...

Page 15: ...the enter key to proceed the calibration process After that the results are shown on display All installed optics must be ok In case of differences refer to chapter 5 Error messages Press enter key t...

Page 16: ...ion the cursor with arrow key up down to select the test to run Press enter key to confirm the first test dialogue is shown on display It is also possible to run a test on a specific reading cell E g...

Page 17: ...position ANL STD to STD ANL then you will be requested to define up to 14 points After measuring results are shown on display together with K and Q values and the regression coefficient R These values...

Page 18: ...nal value SMP Timer definition of the incubation time up to 15 minutes for an activation prior reading of the sample To change this time press the enter key and successively the up arrow key and the d...

Page 19: ...7 3 Auto start dialogue In this mode it is possible to insert the time of automatic switching on of the FOODLABfat instrument if it is connected to the mains and is ready for testing after 45 minutes...

Page 20: ...4 7 5 Info dialogue After selection the following information is shown Registration represents the registration of the instrument Firmware represents the version of programme installed Operator operat...

Page 21: ...own and then switch it on again If the problem persist please contact the technical assistance department for repairing In the following are summarized the instructions to be followed in case specific...

Page 22: ...2 Rif key has been selected 6 Maintenance 6 1 Recommended actions and precautions The FOODLABfat instrument is easy to use and it is normally maintenance free Do not spill any liquid over the instrum...

Page 23: ...your distributor 7 2 Spare parts CDR AEP 119 1 x roll of thermal printer paper 57 mm diam 45 mm CDR ALP 004 external power adapter 100 230V AC to 12V dc CDR 225 058 protective cover for reading cells...

Page 24: ...Operator s manual October 2012 24 7 3 Declaration of conformity...

Page 25: ...d 4 x 3 positions Cuvettes Volume at triangular mark 1000 l mixed polymer for optical use Cups Polystyrol made Sample volume depending on test Throughput depending on test Processor MC 68331 micro pro...

Page 26: ...READ_CHECK chk cc check reading cells cc OK KO DATE date in the days month years format TIME time in the hours minutes format CALIBRATION calibration DATE date in the days months years format TIME tim...

Page 27: ...he rear enable to connect the analyser to a computer FOODLABfat can be connected to LIMS Laboratory Information Management System and the XML Extensible Markup Language protocol 7 6 Waste disposal The...

Page 28: ...tes of particular interest protected by the norms in force ATTENTION Do not hesitate to contact your distributor about any question that is inherent to the demolition procedures of the instrument and...

Page 29: ...Operator s manual October 2012 29...

Reviews: