background image

©

 2019 CDI Meters, Inc. 

10.19 

P

ULSE 

O

UTPUT

 

The pulse output is open-collector, 

that is, a transistor switch to the meter’s 

negative supply. To use it, connect it to the input of the receiving device; also 
connect that input through a 10K Ohm resistor to a suitable positive supply and 
connect the negative supply of the receiving device to the negative supply of the 
meter. If an isolated relay contact is required, install the CDI 5200-IPO isolated 
pulse output and wire it to the receiving device. 

S

ERIAL 

C

OMMUNICATION 

O

PTION

 

The RS-485 signal used with the Serial Communication option is wired to the 
Serial Communication display board. The two threaded openings in the meter 
enclosure permit the signal to be wired from meter to meter without external 
splices. Please refer to the data sheet for the Serial Communication option. 

Using the Outputs 

The milliamp output is scaled so that four milliamps corresponds to zero flow and 
20 milliamps corresponds to the range of the meter indicated on the data sheet. 
The meter will display the milliamp full-scale value for a few seconds on startup. 
It will also display it if the button on the main (lower) circuit board is pressed 
twice. 

The pulse output generates a square wave signal, sending one pulse for each 
cubic foot of air that passes through the meter. The LED blinks with the pulse 
output. At zero flow it may be on or off. The pulse output can be configured as a 
threshold output, and the scaling of both the pulse output and the milliamp output 
can be changed, using the optional configuration cable. 

Using the Display 

Pressing a button indicated by a circle on the front of the meter cycles the display 
through four display options: rate, daily usage, cumulative usage and units of 
measure, the last indicated by a digit: (0) for scfm, (1) for m

3

/min and (2) for 

m

3

/hr. The cumulative values are indicated in thousands of cubic feet or cubic 

meters. Holding the button pressed in either of the cumulative modes (daily 
usage or cumulative usage) will reset that value to zero. The display normally 
defaults to the rate display, but this can be changed, as can the units of measure 
and the decimal point location in the cumulative modes. Please refer to 
information on configuring the display. 

Maintenance 

If oil or dirt accumulates on the probes, the meter will read inaccurately. For this 
reason, we recommend cleaning the probes from time to time. To clean the 
probes, wipe them with a cloth dampened with alcohol or a similar degreaser. It 
may be found that the system is clean enough that cleaning is not needed.  

 

Summary of Contents for CDI 5100

Page 1: ...the plant floor The meters may when necessary be installed outdoors For best visibility avoid direct sunlight Orienting the Display The meter must be installed with flow in the direction indicated by the flow arrow If doing so will cause the display to be upside down remove the cover of the meter lift out the display rotate both 180 and re install Wiring the Meter Wiring must be in accordance with...

Page 2: ...f the cable is 22 gauge or heavier and the distance is no greater than 200 feet CDI 5100 flowmeters require summing remote displays with Rev 24 or later firmware MILLIAMP OUTPUT The 4 20 milliamp output is optically isolated within the meter and it is provided with a factory installed jumper which allows it to source a milliamp signal powered from the meter The resistance of the loop connected to ...

Page 3: ...wice The pulse output generates a square wave signal sending one pulse for each cubic foot of air that passes through the meter The LED blinks with the pulse output At zero flow it may be on or off The pulse output can be configured as a threshold output and the scaling of both the pulse output and the milliamp output can be changed using the optional configuration cable Using the Display Pressing...

Page 4: ...air or replace at its option any defective product at no cost to the buyer The foregoing warranty is in lieu of any other warranty express or implied written or oral including any warranty of merchantability or fitness for a particular purpose CDI s liability arising out of the manufacture sale or supplying of the flowmeter whether based on warranty contract tort or otherwise shall not exceed the ...

Reviews: