PR-T Service Manual
13 8/31/2015
will blink. It means a division is “0.005 (calibration unit, kg or lb)”
(13) Input a division by using [ZERO] key.
(14) Press [TARE] key, and then “dual “ message blinks.“1” will be shown. The third digit,”1”
will blink. It means a dual interval is disable. (0 : disable, 1 : enable)”
(15) Input a dual interval enable by using [ZERO] key.
(16) Press [TARE] key to save the calibration factor, and “C-10 ” message will be shown.
Summary of Contents for PR-T Series
Page 1: ...PR T Service Manual 1 8 31 2015 PR T SERVICE MANUAL...
Page 5: ...PR T Service Manual 5 8 31 2015 1 4 Dimension...
Page 7: ...PR T Service Manual 7 8 31 2015 1 6 SealingMethod...
Page 14: ...PR T Service Manual 14 8 31 2015...
Page 17: ...PR T Service Manual 17 8 31 2015 4 2 Circuit Diagram 4 2 1 Main...
Page 18: ...PR T Service Manual 18 8 31 2015 4 2 2 Power Part...
Page 19: ...PR T Service Manual 19 8 31 2015 4 2 3 Key Part...
Page 20: ...PR T Service Manual 20 8 31 2015 4 2 4 Display Part...
Page 21: ...PR T Service Manual 21 8 31 2015 5 Exploded View 5 1 BODY...
Page 22: ...PR T Service Manual 22 8 31 2015 5 2 Load Cell drawing...
Page 23: ...PR T Service Manual 23 8 31 2015 6 Part 6 1 Part Location Main PCB Top...
Page 24: ...PR T Service Manual 24 8 31 2015 6 1 1 Main PCB Bottom...
Page 25: ...PR T Service Manual 25 8 31 2015 6 1 2 Front Display PCB Top...
Page 26: ...PR T Service Manual 26 8 31 2015 6 1 3 Front Display PCB Bottom...
Page 27: ...PR T Service Manual 27 8 31 2015 6 1 4 Rear Display PCB Top 6 1 5 Rear Display PCB Bottom...
Page 28: ...PR T Service Manual 28 8 31 2015 6 1 6 Pole Display PCB Top...
Page 29: ...PR T Service Manual 29 8 31 2015 6 1 7 Pole Display PCB Bottom...