ER-II
Service Manual
Service Manual
Service Manual
Service Manual
16
7/4/2007
KG/LB
27
TEST
28
HOLD
29
30
NO FUNCTION
31
EURO
60
PERSENT TARE
62
TARE SAVE
63
BOTH SAVE
64
PERSET
65
MUL ”X”
66
PLU
32~59
2.1.7.
Percent Calibration (C-7)
(1) Pressing and holding
“
Calibration Switch
”
press [POWER] key.
After
“
CAL
”
message blinks three times and shows the version of scale, it displays
“
CAL 1
”
message.
(2) Press [ZERO] to display
“
CAL-7
”
.
(3) Press [TARE] key and then it displays
“
per 0
”
message. Select the percent value using the
[numeric] key. You can choose 10~90 percent.
(4) Press [TARE] key and then it displays
“
zero
”
message
(5) Press [TARE] key and then it displays
“
pspan
”
message
(6) Load choice percentage weight of full weight on the platform
(7) Press [TARE] key and then it displays
“
CAL 7
”
message
2.1.8.
Battery Calibration (C-8)
(1) Pressing and holding
“
Calibration Switch
”
press [POWER] key.
After
“
CAL
”
message blinks three times and shows the version of scale, it displays
“
CAL 1
”
message.
(2) Press [ZERO] to display
“
CAL-8
”
.
KEYS
MATRIX KEY CODES
SOFT KEY CODES
“0” through “9”
0 through 9
0 through 9
“C”
10
10
“ON/OFF”
12
12
“ZERO”
13
13
“TARE”
14
14