ER-II
Service Manual
Service Manual
Service Manual
Service Manual
10
7/4/2007
2.
Calibration
2.1.
General Calibration
Pressing and holding calibration switch press [POWER] key to go to calibration mode.
User can move to other mode by using [ZERO] key in the calibration mode.
User also moves to other sub-modes for each mode by using [TARE] key.
Please simply follow below procedure to move to other mode.
(1)
Calibration Mode: Pressing and holding “Calibration Switch” press [POWER] key.
(2)
It displays “CAL-0” after “CAL”, and it blinks the version of scale three times.
(3)
Selecting menu: press [TARE].
(4)
ENTER(Setting) : [TARE] key
MODE
Function
CAL 1
Display normalized AD
CAL 2
Display Keypad infomation-
CAL 3
Weight Setting Mode
“UnLoad”
→
[TARE]
→
“MIDD”
→
[TARE] after loading for 1/3 weight
→
“FULL”
→
[TARE] after loading for Full weight
→
“MIDD”
→
[TARE] after loading for 1/3 weight
→
“END”
CAL 4
Option Setting ( Table 1
참조
)
CAL 5
Display filtered Raw AD
CAL 6
Function setting on each Key ( Table 2
참조
)
CAL 7
% Calibration
CAL 8
Battery calibration
CAL 9
Gravity constant
CAL 10
Set calibration factor
“Unit”
[TARE]
select 0, 1 (0:kg, 1: lb)
[TARE]
“CAPA”
[TARE]
select capacity
[TARE]
“MCAPA”
[TARE]
select mid-capacity
[TARE]
“W-dP”
[TARE]
Select Decimal Point
[TARE]
“ 1 d ”
[TARE]
Select division
[TARE]
“Dual”
[TARE]
Enable dual interval (0:disable, 1:enable)
TARE
CAL 11
Set nation(00 : OIML , 01 : NTEP , 02: KOREA)