7800-25 Series Operations Manual
Rev B 11-10-2020
Page 53 of 56
TROUBLESHOOTING
Problem -- Unit will not turn on
Is the unit connected to a dedicated electrical circuit as defined in the installation
section of the manual?
Is there power at the electric outlet the unit is plugged into?
Is the unit’s power switch turned on?
Problem – The chemical indicator strip did not turn from blue to pink after
sterilization cycle was complete.
H2O2 canister present in H2O2 Canister Housing?
H2O2 canister fluid level full?
H2O2 canister outdated/ expired check date code on side?.
H2O2 heater failed, check Maintenance screen?
Problem – The interlock activates while the door is open (or not fully closed)
Physically push ‘up’ on the lock plunger returning it to the ‘unlocked’ position.
Close
the door
and re-start the sterilization cycle.
Problem – The interlock stays locked after sterilization cycle is complete
Physically push ‘up’ on the lock plunger returning it to the ‘unlocked’ position.
This
will require a flat screwdriver to be inserted through the front side of the lock housing.