7800-25 Series Operations Manual
Rev B 11-10-2020
Page 12 of 56
The unit requires a dedicated electrical connection. Power requirements vary depending
upon the chamber model, see Connecting Electrical Power section. If you have
questions contact Caron Service Department at
Preliminary Cleaning
Your new 7800 H2O2 Sterilization chamber was thoroughly cleaned prior to leaving the
factory. It is recommended, however, to
clean all interior surfaces with a general
purpose laboratory cleaning agent to
remove any shipping dust or dirt prior to using the
product.
Contact Caron if there is any doubt of the compatibility of the cleaning agent being
used with the chamber. After cleaning, dry all interior components with a sterile cloth
as necessary.
Installing the Shelves
Each
new 7800-25 Series H2O2 Sterilization chamber includes
4 wire stainless steel
shelves. Each shelf requires two shelf channels for installation. The left and right shelf
tracks are the same. Prior to installation, take time to consider what the size of the
product being placed in the chamber will be and set the shelf spacing accordingly.
Additional shelving can be purchased through CARON customer service if necessary.
The chamber should be empty when being moved.
To install the shelf channels insert the rear tab on the shelf channel into the rear wall on
the side wall of the chamber. Then insert the front tab into the front pilaster. Push the
entire shelf channel towards the rear of the unit and snap it down into place.
Shelf capacity: each wire shelf is capable of supporting a
uniformly distributed load of 20 lbs, maximum chamber
capacity is 500 pounds (stationary).
Optional shelving is available for supporting heavy loads up to
50 or 150lbs.
Chamber should be empty when being moved.
Do not have multiple loaded shelves out simultaneously or
the chamber may tip.