7800-25 Series Operations Manual
Rev B 11-10-2020
Page 25 of 56
The following screen appears giving the option to cancel or continue with the cycle.
If the
(Cancel) button is pressed a 2
nd
time, then the Sterilization Cycle will
be aborted. The system will stop adding H2O2 in the chamber.
The Sterilization Cycle immediately goes to the Inactivation Stage removing any
residual H2O2 that may be in the chamber before the Sterilization Cyle was aborted.
The
Info button appears in the Status Bar indicating that the the Sterilization
Cycle was manually aborted.
Info Button