6.3  Check the air tubes (leak in tube or inner cells) 

 

 

 

 

- potential leak in the foot section

 

 

 

6.4  Pressure target value invalid

 

 

 

6.5  SD card operating error

   

 

 

 

6.6  Scheduled maintenance notices

 

 

 

6.8 Battery operating error

 

 

 

6.9  Battery charge falls low

 

 

 

6.10  Device internal error

6.1 Pressure sensor function error

The LED row will light up as shown in the 

figure. The LED for the information signal 

reset button blinks and the display indi-

cates failure.

To acknowledge the failure indication 

signal, press the information signal re-

set button. Only the audible part of the 

information signal is acknowledged. The 

visual information signals will remain on 

and the device will not resume previous 

operation.

Immediately contact Carital

®

 service.

Pressure sensor

failure; contact 

service

Määräaikaishuollon 

aikaraja ylittynyt; 

ota yhteys huoltoon

Akun toimintavirhe; 

ota yhteys huoltoon

Akun varaus erittäin 

alhainen; kytke virta-

kaapeli. Sammutukseen 

15:00

Laitteen sisäinen 

virhe; käynnistä 

laite uudelleen

Tarkista ilmaletkut; 

katso irrotettava 

pikaohje

Paineen tavoitearvo 

virheellinen; ota 

yhteys huoltoon

SD-kortin toiminta-

virhe; ota yhteys 

huoltoon

Summary of Contents for Optima7080

Page 1: ...KOHJ H 01 EN1 1 7 15112021 1 User s guide Optima7080 Optima8590...

Page 2: ...r 12 3 4 Rehab add on for the cover 13 3 5 Comfort add on for the Medicase hygiene cover 13 3 6 Taking off the cover 14 3 7 Putting on the cover 15 4 Commissioning 17 4 1 Components of the mattress sy...

Page 3: ...7 1 Cleaning 39 7 1 1 Controller and tube system 39 7 1 2 Cells 39 7 1 3 Medicase Mediresc and antistatic Medicase cover 39 7 1 4 Comfort add on for the cover 40 7 1 5 Rehab add on for the cover 40 7...

Page 4: ...745 MDR Main fuse infor mation Keep protected from rain Fragile handle with care Do not use if the packaging is damaged or has been opened Store away from heat See user s guide Type BF device Permitte...

Page 5: ...signals and the adjustment of different adjustment regions Operating modes Normal operation Sitting operation Other NFC tag location in device note this feature is not yet implemented Keypad lock Bat...

Page 6: ...ital Optima mattress system has two 2 different variants which are deter mined by the width of the cells Optima7080 the identifier OP7 at the beginning of the REF field on the name plate of the contro...

Page 7: ...mattress system independently Keep this guide 2 7 Warnings This guide applies only to second generation Carital controllers A second genera tion controller can be identified with its serial number beg...

Page 8: ...of the cells by checking the REF field on the controller and the cell ID label Ensure that the settings of the device do not change unintentionally for example because of children or pets If necessar...

Page 9: ...cover make sure the patient is always lying on the cells not on the side supports The side handles of the Mediresc cover are intended only for general handling of the mattress without patient load Fai...

Page 10: ...ther electric devices or in a stacked configuration as this may interfere with the controller s operation If the above use is necessary ensure the normal operation of the controller by monitoring it U...

Page 11: ...ice instead The user is responsible for any and all consequences resulting from neglecting ser vice Scheduled maintenance may only be carried out by Carital service 3 Covers This section presents the...

Page 12: ...cover at the foot end In the Mediresc rescue cover the controller s power cable is routed through a conduit integrated to the bottom of the mattress accessible by a zipper preventing clamping of the p...

Page 13: ...e may be clamped by the sides or folding parts of the bed 3 4 Rehab add on for the cover All covers are available with integrated Rehab side bars The side support bars improve the support of the suppo...

Page 14: ...the zipper fully to expose the air tube system and power cable 5 Open the mattress zipper from the end until the air tube connectors are visible 2 Start by pulling down the zipper on the tube sleeve...

Page 15: ...eck that the necessary accessories are readily available power cable air tubes suspension straps 2 and side support bars only covers with the Rehab add on 2 If you notice any protruding inner cells ev...

Page 16: ...cell air tubes out of their hole Note If the cover includes the Rehab add on place the side support bars as instructed in section 3 4 9 Close the zipper on the tube sleeve 6 Pass the strap through th...

Page 17: ...e mattress system Controller Controller Type plate stickers on the side and at the bot tom of the controller contain the device iden tification information The lockable power cable 5 m The power cable...

Page 18: ...the device and provides an example of troubleshooting The double sided quick guide is found on the back of the controller and is released for viewing by raising it upwards Be sure to put the quick gui...

Page 19: ...gaps in the lying surface an extra bottom support foam optional accessory can be used inside the cover underneath the cells The support surface should fit the bed so that it does not extend over the e...

Page 20: ...o the moving back part of the lying surface not to the fixed part of the bed body Do not use extra bed sheets pillows or heavy positioning pillows on the mattress If side rails are used with the Carit...

Page 21: ...tube sleeve is as secure as possible in relation to the bed structure 4 Connect the power cable to the mains con nector of the controller 2 To connect the three colour coded tubes of the air tube sys...

Page 22: ...er using two hands on both sides of the body or according to the instructions with the attached suspender 6 Pass the protective tube sleeve as far up as possible towards the controller Close the zippe...

Page 23: ...ube sleeve has been routed in the bed structure in such a way that it cannot be clamped by the folding parts of the bed Ensure that the label indicating the correct direction of installation of the ma...

Page 24: ...size of the patient such that the pressure in all the adjustment sections are optimally placed to be regulated by the controller in correspond ence to the patient s body parts Make sure that the cont...

Page 25: ...f the controller To turn off the controller in any operating mode press the device s standby button The controller can be turned off in any operating mode by pressing the de vice s standby button The...

Page 26: ...tch to Sitting oper ation The device will first adjust the torso then the foot region and finally the head The Sitting operation LED is lit and the cell LEDs will move up or down The device has activa...

Page 27: ...hanged This allows the controller to monitor the pressure in the torso cells If the pressure values of the torso cells are continuously outside the permissible limit values for 45 minutes the device w...

Page 28: ...s function With the device in operation press the Supine operation button to activate the function The device activates Supine operation The device will first adjust the torso then the foot region and...

Page 29: ...Once the patient transport is finished connect the controller back to the mains by connecting the power cable to an electrical outlet The cells will not deflate during transport The controller must be...

Page 30: ...he cell s tube system and the power cable from the controller turn off the controller and contact Carital service 6 Information signals If the controller detects a failure or wishes to inform the user...

Page 31: ...nowledge the failure indication signal press the information signal re set button Only the audible part of the information signal is acknowledged The visual information signals will remain on and the...

Page 32: ...er 6 3 Check the air tubes leak in tube or inner cells The information signal reset button blinks and the display reads Check the air tubes See the removable quick guide This information signal is dis...

Page 33: ...e Close the tube sleeve When you have checked the above ac knowledge the information signal by pressing and holding down the informa tion signal reset button for three sec onds If the information sign...

Page 34: ...limit for scheduled maintenance is approaching prepare for maintenance Prepare to send the controller in for scheduled maintenance after one 1 month After this the device will display a re minder for...

Page 35: ...hown in an illogical way Turn the controller off and on with the standby button shown in the figure The device will resume Normal operation af ter restarting If the device does not resume Normal op er...

Page 36: ...t resume previous operation Immediately contact Carital service 6 9 Battery charge falls low When the charge of the internal battery falls to a very low level 7 2V 7 0V the controller will provide an...

Page 37: ...isconnected from the mains before an ade quate charge level has been reached the device will shut down itself after 5 seconds after booting To maintain battery performance connect the controller to AC...

Page 38: ...by a party other than Carital service If the mattress system behaves contrary to the functions and situa tions described in this user s manual disconnect the air tubing from the cell s tube system an...

Page 39: ...sed to urea 7 1 1 Controller and tube system Disinfect by wiping using regular cleaning and disinfection agents including ethanol solu tions 60 80 chlorine solutions max 1 000 ppm Dry at room temperat...

Page 40: ...y Do not iron Do not dry wash Do not use conditioners Ensure that the cover is entirely dry before commissioning it 7 1 5 Rehab add on for the cover Remove the integrated foam support from the cover A...

Page 41: ...e 7 2 3 Cells The condition of cells must be inspected When cleaning If you suspect that the cover is broken or the interior is contaminated When the patient changes or weekly in long term care Strip...

Page 42: ...violation of the instructions specified in the user s guide or it is not cleaned of body secretions containing urea in particular or the mattress system is used by a prominently sweating or mobile pa...

Page 43: ...Service To change from one tab to another in the maintenance view press the information signal reset button To return from the maintenance view to the normal operating mode press and hold down the in...

Page 44: ...o 50 C 35 C to 70 C with vapour pressure 50 hPa Air humidity max 90 Store in a clean and dry place The cells and the cover can be rolled up for storage using for example the trans port bag accessory A...

Page 45: ...can be recycled with cardboard The styrofoam packaging supports and packaging plastic can be recycled with plastic packaging 9 Warranty The Carital Optima mattress system has three year warranty 36 mo...

Page 46: ...Nominal input power max 35W Battery type Lithium ion 7 26V capacity 2 650mAh manufacturer Cell tech Oy Varta Storage GmbH Non rechargeable battery type CR2032 lithium ion 3 0V capacity 230mAh manufact...

Page 47: ...014 IEC 60601 1 6 2010 IEC 60601 1 6 2010 AMD1 2013 IEC 60601 1 11 2015 IEC 62304 2006 IEC 62304 2006 AMD1 2015 IEC 62366 2007 IEC 62366 2007 AMD1 2014 IEC 60601 2 52 2009 subclause 201 9 101 EN ISO 1...

Page 48: ...1 EN1 1 7 15112021 11 Contact details of the manufacturer and service Manufactured by MediMattress Ltd Haukilahdenkatu 4 FI 00550 HELSINKI tel 358 306 40 40 40 info medimattress fi Service add local s...

Page 49: ...phase at 0 0 UT 250 300 cycle 0 UT 0 5 cycle at 0 45 90 135 180 225 270 and 315 0 UT 1 cycle 70 UT 25 30 cycles Single phase at 0 0 UT 250 300 cycle Mains power quality should be that of a typical com...

Reviews: