4.5. Things to check before use
When attaching the mattress system to the
bed structure, ensure that both suspension
straps are attached and tightened according
to instructions.
Ensure that the controller has been suspended
in as central position as possible at the foot
end of the bed. Ensure that there is sufficient
space around the controller for operation and
unobstructed disconnection.
Make sure the tube sleeve has been routed in
the bed structure in such a way that it cannot
be clamped by the folding parts of the bed.
Ensure that the label indicating the correct
direction of installation of the mattress sys-
tem is located at the foot end and facing up-
ward.
TÄMÄ PUOLI YLÖS
THIS SIDE UP
JALKOPÄÄ
FOOT END
IET-G-02-1.2
To guarantee fault-free operation of the controller, the power cable
must always be connected to the mains, with the exception of short
patient transports and power outages.
Always place the controller in such a way that it can easily be dis-
connected from the mains and that the controller’s switches are
unobstructed.