For instructions on what to do if battery charge falls below the level required for
normal operation, see section 6.9 (Information signals - Battery charge falls low).
5.9 Resuscitation situations
WHEN RESUSCITATING:
Turn off the device from the standby button and begin CPR immediately
without deflating the cells.
When resuscitating, turn off the device from the standby button
and start CPR immediately without deflating the cells. Do not use
Firm
operation when resuscitating.
5.10 Fault situations
The identified fault situations of the mattress system and detecting them is described in
the sections 6. and 7.2.
If the mattress system behaves contrary to the functions and situa-
tions described in this user’s manual, disconnect the air tubing from
the cell’s tube system and the power cable from the controller, turn
off the controller, and contact Carital
®
service.
6 Information signals
If the controller detects a failure or wishes to inform the user, it will provide audible
and visual indication using the display, information signal LED and LED bars. This section
describes how the information signals must be interpreted and what action they require
from users.
Below is a list of information signals in the LED display and references to more detailed
troubleshooting guides:
6.1 Pressure sensor function error
6.2 Elevated pressure ulcer risk (sitting operation)
6.3 Check the air tubes (leak in tube or inner cells)
- potential leak in the torso section
6.3 Check the air tubes (leak in tube or inner cells)
- potential leak in the head section
Paineanturin toimin-
tavirhe; ota yhteys
huoltoon
Tarkista ilmaletkut;
katso irrotettava
pikaohje
Kohonnut painehaa-
variski; tarkista
toiminto
Tarkista ilmaletkut;
katso irrotettava
pikaohje