The device activates
Firm
operation.
The device will first adjust the torso,
then the foot region and finally the head.
The cell LEDs will go up.
The device has activated
Firm
operation.
The
Firm
operation LED is lit and the cell
LEDs are lit on the centre row.
5.5 Silent operation
The Silent
operation can be selected in a situation where you want to minimize the op
-
eration of your device's control cycles and pump function, for example during night time
when the patient is asleep.
The controller adjusts the cell pressures to
Normal
operation pre-set values and then
retains them unchanged. This allows the controller to monitor the pressure in the torso
cells. If the pressure values of the torso cells are continuously outside the permissible
limit values for 45 minutes, the device will start adjusting all the cell parts again, start-
ing from the torso cells.
Eight (8) hours after selecting the
Silent
operation, the device automatically returns to
Normal
operation.
With the device in operation, press the
Silent
operation button to activate the
function.
The device activates
Silent
operation.
The device will first adjust the torso,
then the foot region and finally the head.
The operation LED is lit and the cell LEDs
will go up or down.
The device has activated
Silent
opera-
tion. The operation LED is lit and the cell
LEDs are lit on the centre row.
- Firm -
Adjusting
- Firm -
OK
- Silent-
Adjusting
- Silent-
OK