OPERATION MANUAL

The battery failure will be triggered 
when orange LED is flashing with 
audible indicator. The battery is not 
user replaceable, please contact your 
authorized local distributor for assistance. 

Multi-Cycle Time Button   
Press the button to set up the alternating 
cycle time as 10, or 15, or 20 minutes to 
meet a variety of patient’s requirements.

Low Pressure Indicator   
Both continuous flashing of orange LED 
and audible indicator will be triggered to 
notify medical staff when the cushion has 
insufficient internal pressure. The power 
unit will automatically turn off if the low 
pressure indicator stays for 15 minutes. 

Panel Lock/ Unlock  
The power unit locks automatically when 
the function buttons are not touched for 
3 minutes or so. All function buttons 
are locked out. Simply press        for 3 
seconds to unlock the power unit and 
panel. 

Seating Cushion Mode
Press 

 and      simultaneously 

for at least 3 seconds, the display will 
change from 

 to 

 which 

indicates the device has already switched 

Side Panel Features (Figure 5b)

Rear side Panel Features (Figure 5c)

Control Panel Features (Figure 5a)

1

2

3

4

5

Summary of Contents for Cari Chair

Page 1: ...Cari Chair Operation Manual Anti Decubitus Air Alternating Pressure Cushion Replacement System...

Page 2: ...ESE INSTRUCTIONS CAREFULLY BEFORE USE AND OBSERVE THE SAFETY INSTRUCTIONS AND THE REQUIREMENTS FOR THE OPERATION AND MAINTENANCE OF THE DEVICE Use genuine Carilex components are essential for optimal...

Page 3: ...stem Package 5 5 0 Features 6 6 0 Mattress and Power Unit Installations 8 7 0 Program Settings 8 8 0 Cleaning Instruction 9 Section Description Page 9 0 Routine Maintenance 9 10 0 Troubleshooting 10 1...

Page 4: ...of pressure ulcer This device is intended to prevent pressure ulcers by facilitating blood circulation and decreasing pressure of each tissue s contact area If there s any question please always consu...

Page 5: ...ients to avoid of cross contamination 8 Be sure to verify the patient weight does not exceed system weight capacity The maximum weight capacity of this device is 113 kg 250 lb 9 Charging the power uni...

Page 6: ...to ensure maximum pressure relief The yellow LED indicates the device is monitoring the cushion s internal pressure while the compressor is inflating Battery Indicator The green LED bar indicates the...

Page 7: ...will be triggered to notify medical staff when the cushion has insufficient internal pressure The power unit will automatically turn off if the low pressure indicator stays for 15 minutes Panel Lock...

Page 8: ...ng to the next step 3 Secure the power unit to the rear of the chair by using dedicated carry bag 4 Firmly connect the air hose couplings to the quick release couplers on the power unit s air outlet 5...

Page 9: ...lor might run insignificantly 4 DO NOT use electric or tumble dryers 5 DO NOT iron WARNING Always unplug the power unit before cleaning Routine cleaning of power unit can be done by wiping down with d...

Page 10: ...rmly into place Verify if patient weight setting is correct Increase or decrease weight setting until appropriate pressure is reached Inspect air filter for dust Clean or replace air filter Lift cushi...

Page 11: ...roduct on which the production lot number has been removed or defaced on which repair has been attempted by any person or agency not authorized by Carilex or if in the sole opinion of Carilex that the...

Page 12: ...n 180 x 50 x 14 cm 26 air cells 70D nylon with TPU lamination Coverlet Two way stretch 40 polyurethane 60 polyester Wheel Chair Cushion Set Cushion dimension 17 7 x 17 7 x 3 5 in 45 x 45 x 9 cm 10 air...

Page 13: ...ir Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast transient burst IEC 61000 4 4 2kV for power...

Page 14: ...ress m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vic...

Reviews: