ANTI-DECUBITUS AIR ALTERNATING PRESSURE CUSHION REPLACEMENT SYSTEM

5

English

Comfort Level Button   
Simply press this button to adjust the 
patient sensation from 1 to 5 according to 
each individual need. The scale is only an 
approximation. Please adjust the comfort 
level when patient feels the cushion is too 

soft or too firm. Caregivers should always 

perform a hand check by placing their 
hands underneath patient’s pelvis area to 
check if there is sufficient air support to 
ensure the patient is not bottoming out.

Static Mode Button   
Press this button then the green LED lights 
up and the Multi-Cycle Time displays 0 
means the device is in Static Mode. The 
pressure re-distribution function provides 
optimal internal pressure for each different 
comfort level. Simply press          to return 
the device to Alternating Mode.

Pressure Monitoring Indicator   
Power unit features an integrated pressure 
sensor which can monitor cushion’s 
internal pressure 24 hours to achieve 
optimal internal pressure and to ensure 
maximum pressure relief. The yellow 
LED indicates the device is monitoring 
the cushion’s internal pressure while the 

compressor is inflating. 

Battery Indicator   
The green LED bar indicates the battery 
level or charging status. If the power unit 
is low on power, the orange LED will 
light up to indicate that the power unit 
needs to charge for 4 hours to battery full. 

CAUTION: 

ENSURE THAT THERE ARE  NO 
PROTRUDING OBJECTS, SHARP 
POINTS OR CHAIR SPRINGS UNDER 
THE CUSHION AS THESE COULD 
PUNCTURE THE AIR CELLS AND 
AFFECT THE PERFORMANCE.

4.0 System Package

Power Unit Package

* Power unit x 1
* Power Adapter x 1
* Operation Manual x 1

Geri Chair Cushion Package

* Geri Chair Cushion with Coverlet x 1

MobiCare Cushion Package

* MobiCare Cushion with Coverlet x 1

Wheelchair Cushion Package

* Wheelchair Cushion with Coverlet x 1

Optional Accessory

* Carry Bag
* Bed Hanger
* IV Pole Hanger

5.0 Features

Control Panel Features

Figure 5a on page 6

Power/ Mute/ Unlock Button  
[1] Press this button to turn on the power 

unit and light up the green LED. The 

screen displays 

 to indicate 

the device is in default Geri Chair 

mode. 

[2] When audible indicator is sounding, 

pressed this button to mute the audible 

indicator.

 

Summary of Contents for Cari Chair

Page 1: ...Cari Chair Operation Manual Anti Decubitus Air Alternating Pressure Cushion Replacement System...

Page 2: ...ESE INSTRUCTIONS CAREFULLY BEFORE USE AND OBSERVE THE SAFETY INSTRUCTIONS AND THE REQUIREMENTS FOR THE OPERATION AND MAINTENANCE OF THE DEVICE Use genuine Carilex components are essential for optimal...

Page 3: ...stem Package 5 5 0 Features 6 6 0 Mattress and Power Unit Installations 8 7 0 Program Settings 8 8 0 Cleaning Instruction 9 Section Description Page 9 0 Routine Maintenance 9 10 0 Troubleshooting 10 1...

Page 4: ...of pressure ulcer This device is intended to prevent pressure ulcers by facilitating blood circulation and decreasing pressure of each tissue s contact area If there s any question please always consu...

Page 5: ...ients to avoid of cross contamination 8 Be sure to verify the patient weight does not exceed system weight capacity The maximum weight capacity of this device is 113 kg 250 lb 9 Charging the power uni...

Page 6: ...to ensure maximum pressure relief The yellow LED indicates the device is monitoring the cushion s internal pressure while the compressor is inflating Battery Indicator The green LED bar indicates the...

Page 7: ...will be triggered to notify medical staff when the cushion has insufficient internal pressure The power unit will automatically turn off if the low pressure indicator stays for 15 minutes Panel Lock...

Page 8: ...ng to the next step 3 Secure the power unit to the rear of the chair by using dedicated carry bag 4 Firmly connect the air hose couplings to the quick release couplers on the power unit s air outlet 5...

Page 9: ...lor might run insignificantly 4 DO NOT use electric or tumble dryers 5 DO NOT iron WARNING Always unplug the power unit before cleaning Routine cleaning of power unit can be done by wiping down with d...

Page 10: ...rmly into place Verify if patient weight setting is correct Increase or decrease weight setting until appropriate pressure is reached Inspect air filter for dust Clean or replace air filter Lift cushi...

Page 11: ...roduct on which the production lot number has been removed or defaced on which repair has been attempted by any person or agency not authorized by Carilex or if in the sole opinion of Carilex that the...

Page 12: ...n 180 x 50 x 14 cm 26 air cells 70D nylon with TPU lamination Coverlet Two way stretch 40 polyurethane 60 polyester Wheel Chair Cushion Set Cushion dimension 17 7 x 17 7 x 3 5 in 45 x 45 x 9 cm 10 air...

Page 13: ...ir Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast transient burst IEC 61000 4 4 2kV for power...

Page 14: ...ress m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vic...

Reviews: