14.0 Technical Data

Air flow rate of power unit 

Open flow 2.5 liters/min

Power Input:  

AC 100-240V - 50/60Hz (for Adapter)

Power Output  

DC 9.1V 3.33A or DC 9V 3A (for Adapter)

Power Adaptor  

GlobalTek, Inc. GTM91120-3010.5-1.4-FW
SINPRO HPU32A-104

Power Consumption  

Max: 27W (with charger)
Max.9W (without charger)

Operation mode

Geri Chair Mode & Seating Cushion Mode

Rechargeable battery

Lithium-ion Capacity: 4400mAh 
Charging Time: 4hrs 
Using Time: approx.12 hours
Typical operation time (charge & discharge cycle): 300 
times

Dimensions (WxHxD)

9.6 x 6.1 x 2.4 inch / 24.5 x 15.5 x 6.1 cm

Weight (basic unit)

2.03lb / 0.92kg

Geri Chair Cushion Set :

Cushion dimension : 70.9 x 19.7 x 5.5 in / 180 x 50 x 14 
cm
26 air cells, 70D nylon with TPU lamination.
Coverlet: Two way stretch – 40% polyurethane, 60% 
polyester.

Wheel Chair Cushion Set :

Cushion dimension :17.7 x 17.7 x 3.5 in / 45 x 45 x 9 cm
10 air cells, 210D nylon with TPU lamination.
Coverlet: Two way stretch –40% polyurethane, 60% 
polyester.

MobiCare Cushion Set :

Cushion dimension :17.7 x 17.7 x 3.5 in / 45 x 45 x 9 cm 
(with foam base)
10 air cells, 210D nylon with TPU lamination.
Coverlet: Two way stretch –40% polyurethane, 60% 
polyester.
Cushion foam base: PU foam.

IP22

Against ingress of solid foreign objects 
12.5 mm diameter. 

Against direct sprays of water up to 15˚ from the vertical

Protection class according to IEC60601-1 Class II

Operation Conditions

Temperature range: 5˚C(41˚F) to 40˚C(104˚F) 

Relative Humidity Range: 15%~60% noncondensing.
Atmosphere range: 700hPa-1060hPa

Transport and storage conditions 

Temperature Range: -25˚C (-13˚F) to 70˚C(158˚F) 

Relative Humidity Range: 0%~93% noncondensing

Summary of Contents for Cari Chair

Page 1: ...Cari Chair Operation Manual Anti Decubitus Air Alternating Pressure Cushion Replacement System...

Page 2: ...ESE INSTRUCTIONS CAREFULLY BEFORE USE AND OBSERVE THE SAFETY INSTRUCTIONS AND THE REQUIREMENTS FOR THE OPERATION AND MAINTENANCE OF THE DEVICE Use genuine Carilex components are essential for optimal...

Page 3: ...stem Package 5 5 0 Features 6 6 0 Mattress and Power Unit Installations 8 7 0 Program Settings 8 8 0 Cleaning Instruction 9 Section Description Page 9 0 Routine Maintenance 9 10 0 Troubleshooting 10 1...

Page 4: ...of pressure ulcer This device is intended to prevent pressure ulcers by facilitating blood circulation and decreasing pressure of each tissue s contact area If there s any question please always consu...

Page 5: ...ients to avoid of cross contamination 8 Be sure to verify the patient weight does not exceed system weight capacity The maximum weight capacity of this device is 113 kg 250 lb 9 Charging the power uni...

Page 6: ...to ensure maximum pressure relief The yellow LED indicates the device is monitoring the cushion s internal pressure while the compressor is inflating Battery Indicator The green LED bar indicates the...

Page 7: ...will be triggered to notify medical staff when the cushion has insufficient internal pressure The power unit will automatically turn off if the low pressure indicator stays for 15 minutes Panel Lock...

Page 8: ...ng to the next step 3 Secure the power unit to the rear of the chair by using dedicated carry bag 4 Firmly connect the air hose couplings to the quick release couplers on the power unit s air outlet 5...

Page 9: ...lor might run insignificantly 4 DO NOT use electric or tumble dryers 5 DO NOT iron WARNING Always unplug the power unit before cleaning Routine cleaning of power unit can be done by wiping down with d...

Page 10: ...rmly into place Verify if patient weight setting is correct Increase or decrease weight setting until appropriate pressure is reached Inspect air filter for dust Clean or replace air filter Lift cushi...

Page 11: ...roduct on which the production lot number has been removed or defaced on which repair has been attempted by any person or agency not authorized by Carilex or if in the sole opinion of Carilex that the...

Page 12: ...n 180 x 50 x 14 cm 26 air cells 70D nylon with TPU lamination Coverlet Two way stretch 40 polyurethane 60 polyester Wheel Chair Cushion Set Cushion dimension 17 7 x 17 7 x 3 5 in 45 x 45 x 9 cm 10 air...

Page 13: ...ir Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast transient burst IEC 61000 4 4 2kV for power...

Page 14: ...ress m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vic...

Reviews: