![Carestream DRYVIEW 5950 User Manual Download Page 42](http://html1.mh-extra.com/html/carestream/dryview-5950/dryview-5950_user-manual_2557844042.webp)
Maintenance and Troubleshooting
3–14
AA2204_en | 2012-07-27
Condition Codes
Condition codes are shown on the display screen in the order in which they are
generated. If there is more than one code associated with the current condition of the
imager, the first code is shown on the display screen for six seconds, while other codes in
the list display for three seconds as the list is cycled. You can also view these codes and
messages at the Web Portal.
Display
Screen
Web
Portal
Web Portal
Message
Action
004
01004
MIM Core: Internal
Software Error
Restart the imager.
If the error persists, call for service.
200
04200
MIM Core: Disk Full
Load the requested film type and size for jobs
that are waiting for media.
If the error persists, call for service.
400
06400
MIM Core: Image
Page Error
Resend the print job from the image source.
If the error persists, call for service.
410
06410
MIM Core: Image
Rendering Error
Resend the print job from the image source.
If the error persists, call for service.
411
06411
MIM Core: Image
Data Error
Resend the print job from the image source.
If the error persists, call for service.
420
06420
MIM Core: Internal
Software Error
Resend the print job from the image source.
If the error persists, call for service.
430
06430
MIM Core: Internal
Software Error
Resend the print job from the image source.
If the error persists, call for service.
001
10001
MIS: Internal
Software Error
Restart the imager.
If the error persists, call for service.
003
10003
MIS: Image Buffer
Error
Restart the imager.
If the error persists, call for service.
015
10015
MIS: Database Error
Restart the imager.
If the error persists, call for service.
910
10910
MIS: MCS
Communication
Failure
Restart the imager.
If the error persists, call for service.
006
20006
Disconnected or
faulty network cable
Check and reconnect the network cable on both
ends.
If the error persists, call for service.
SHORT TERM CONFIDENTIAL