Maintenance and Troubleshooting
3–12
AA2204_en | 2012-07-27
Related topics
Calibrate the Imager for the Loaded Film
Film
cartridge is
empty
Power symbol is yellow
Pause and Calibrate
symbols are off
Film count is flashing “0”
Empty
and/or sheet
count of 0
The film cartridge is inserted,
but the sheet count is 0. Insert a
new cartridge.
Manual
mode
Status code: 002
Power symbol is green
Calibrate symbol is yellow
AIQC Off
(with normal
tray
information)
The film in this cartridge does
not meet AIQC standards.
However, the imager prints if
ready.
Invalid film
cartridge
Power symbol is yellow
Pause and Calibrate
symbols are off
Film count is replaced by
dashes
Invalid Film
Tray
There is a film cartridge in the
film supply, but it does not
contain a liner/RF tag. Install a
new film cartridge.
Ready
Power symbol is green
Calibrate symbol is on
Normal Tray
Info
The film cartridge is ready for
use.
Requires
calibration
Power symbol is yellow
Calibrate symbol is on
Requires
Calibration
The film cartridge must be
calibrated before the imager can
print.
Film
cartridge is
not detected
Power symbol is yellow
Pause and Calibrate
symbols are off
Film count is blank
No Film Tray
The film cartridge is not fully
inserted in the imager. Insert the
cartridge.
Not ready
Various
Not Ready
When the conditions are
corrected, the imager can print.
Cartridge
closure is
pending
Power symbol is green
Calibrate symbol is off
Pause symbol is on
Pause
Requested
You pressed the Pause button,
but the rollback has not started
because films are still moving
through the imager. When the
closure is complete, the Pause
symbol is off. At this time, you
can remove the film cartridge.
Film
Cartridge
State
Display Screen
Web Portal
Description
SHORT TERM CONFIDENTIAL