Acquiring a TMJ x4 Image for Adult and Pediatric Patients
5–6
ACQUIRING IMAGES
Acquiring a TMJ x4 Image for Adult and Pediatric
Patients
Before acquiring a TMJ x4 image, check that you have:
•
Reset the unit rotative arm to the start position for patient to enter the unit.
•
Selected the patient record.
•
Accessed the
Imaging window
.
•
Accessed the
Panoramic Acquisition
window.
To acquire a TMJ x4 image, follow these steps:
1. From the
Panoramic Acquisition
window, in the
Program pane
, select the TMJ and
click .
2. Remove
place
on the chin rest support and cover it with a hygienic
sheath.
3. Press and hold
to raise the chin rest to the TMJ x4 positioning indicator
4. Correctly position the patient using the 2 positioning laser beams.
5. Ask the patient to remain still and close their eyes. Acquire an image with the mouth
closed.
6. To acquire an image with the mouth open, on the
Control Panel,
click
to reset
the unit rotative arm.
7.
Ask the patient to stay in the same position and open their mouth. Acquire an image
with the mouth open.
8. Do the following when the acquisition is finished:
•
Open the temple supports and release the patient.
•
Remove the hygienic sheath from the bite block.
•
Reset the unit rotative arm for the next acquisition.
Summary of Contents for CS 9000
Page 1: ...CS 9000 AND CS 9000 3D User Guide...
Page 6: ...Conventions in this Guide 1 2 About This Guide...
Page 28: ...Acquisition Interface 3 12 IMAGING SOFTWARE OVERVIEW...
Page 50: ...X Ray Dose Emission Information 5 20 ACQUIRING IMAGES...
Page 58: ...Controlling the Image Quality 7 6 MAINTENANCE...