page
58
/
62
- Shield patients when possible
- Repeat images only when necessary
•
Tables of Exposure Times for children
Patient
Teeth
Angle of
inclination
SSD: 200mm(8 inch)
kV
mA
sec
Child
DR
Incisor
Maxilla : +45
˚
Mandible : -25
˚
70
3.5
0.10
Canine
Maxilla : +45
˚
Mandible : -20
˚
70
3.5
0.12
Upper
Molar
Maxilla : +30
˚
70
3.5
0.15
Lower
Molar
Mandible : -5
˚
70
3.5
0.14
Bitewing
+5
˚
~ +8
˚
70
3.5
0.19
CR
Incisor
Maxilla : +45
˚
Mandible : -25
˚
70
3.5
0.24
Canine
Maxilla : +45
˚
Mandible : -20
˚
70
3.5
0.30
Upper
Molar
Maxilla : +30
˚
70
3.5
0.39
Lower
Molar
Mandible : -5
˚
70
3.5
0.36
Bitewing
+5
˚
~ +8
˚
70
3.5
0.50
FL
Incisor
Maxilla : +45
˚
Mandible : -25
˚
70
3.5
0.33
Canine
Maxilla : +45
˚
Mandible : -20
˚
70
3.5
0.39
Upper
Molar
Maxilla : +30
˚
70
3.5
0.50
Lower
Molar
Mandible : -5
˚
70
3.5
0.45
Bitewing
+5
˚
~ +8
˚
70
3.5
0.65
Use of equipment and exposure settings designed for an average-sized adult can result in
excessive radiation exposure for a smaller patient, especially pediatric. Pediatric patients
may be more radio-sensitive than adults (i.e., the cancer risk per unit dose of ionizing
radiation is higher), and so unnecessary radiation exposure is of particular concern for
pediatric patients. Please use caution when configuring the Portable Dental X-ray (CS
2400P) by considering the patient’s age, size, body habitus, and clinical indication when
verifying exposure time settings.
For safety for the pediatric patients, In child mode, exposure dose of Portable Dental X-ray