Troubleshooting
Error Code
Potential Problem
Corrective Action
E02
(Tube current
under)
If it is lower than the set tube current
CS 2400P will require authorized
service.
E03
(Generator High
voltage cable
disconnect)
Generator high voltage cable not
connected
CS 2400P will require authorized
service.
E04
(High voltage
under)
If the set tube voltage is low
CS 2400P will require authorized
service.
E11
(Generator
filament cable
disconnect)
If the generator filament cable is not
connected
CS 2400P will require authorized
service.
E62
(Trigger the
exposure switch)
When the exposing button shtus down
abnormally
The “LOC” message is displayed
on the lcd screen and is released
using the light button.
4.15
Charging the Battery
1) Connect the charging cable to Cradle
2) Plug the cradle into an electrical socket via the power supply unit provided.
If using it overseas please ensure that an appropriate adaptor for the country of visit is
purchased prior to departure.
3) LED indicator lights on and turns Orange when provide an electricity.
4) Docking the equipment on the cradle properly.
5) The Super Capacitor is charged when the LED turns Blue.
Do not turn on the equipment during charging.