100 kHz repetition frequency
Surge
IEC 61000-4-5
±
0.5 kV,
±
1.0 kV Line-to-line
±
1.0 kV,
±
2.0 kV Line-to-ground
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips
IEC 61000-4-11
0 %
UT
, 0,5 cycle, At 0°, 45°, 90°,
135°, 180°, 225°, 270° and 315°
0 %
UT
, 1 cycle and 70 %
UT
, 25/30
cycles
Single phase: at 0°
Mains power quality should be
that of a typical commercial or
hospital environment, If the user of
the CS 2400P model device
requires continued operation
during power mains interruptions,
it is recommended that the CS
2400P model device be powered
from an uninterruptible power
supply or a battery
Voltage
interruptions
IEC 61000-4-11
0 %
UT
, 250/300 cycle
RATED power
frequency
magnetic fields
(50 Hz/60 Hz)
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
If X-ray output distortion occurs, it
may be necessary to position the
CS 2400P model device further
from sources of power frequency
magnetic fields or to install
magnetic shielding. The power
frequency magnetic field should
be measured in the intended
installation location to assure that
it is sufficiently low.
Note :
UT
is the a.c. mains voltage prior to application of the test level.