Configuring the MicroDL
The MicroDL must be configured for single patient (asthma
management) or multiple patient use. Multiple patient use is
configured by hand on the instrument itself and single patient use
may be configured either by hand or more conveniently by the SPCS
software provided (see SPCS installation and operation).
When configuring for single patient mode a patient identifier must
be entered. This number will be used to identify data when printing
or uploading results. The age, height and sex of the patient
together with any ethnic correction must also be entered. This data
is used to calculate the predicted values for FEV1, FVC, PEF, and
FEV1/FVC. The ethnic correction reduces the predicted values of the
volume measurements by the percentage entered. Finally, the
levels at which the amber and red warning lights are activated are
entered.
These levels are expressed as a percentage of the predicted values
when the unit is configured by hand but the patient’s personal best
values may be used when the unit is configured using SPCS.
When using the MicroDL in multiple patient mode the patient details
are entered each time the unit is turned on. Therefore, when
configuring the unit for multiple patient use, only the warning light
levels are entered.
To configure the unit, depress the off key, turn the unit on and
release the off key. If there is data currently stored in the unit then
the number of records displayed:
Summary of Contents for MicroDL
Page 1: ...1 MicroDL Operating Manual English Bedienungsanleitung Deutsch...
Page 2: ......
Page 4: ...2 IV 103 104 MicroDL 106 MicroDL SPCS 129 131 131 132 134...
Page 36: ......
Page 40: ...37 MicroDL 1 Fig 1 of enter 3 4 on 2...
Page 73: ...70 MicroDL 1 Fig 1 of enter 3 4 on 2...
Page 106: ......
Page 107: ...103 MicroDL IV MicroDL CareFusion MicroDL Spirometry PC Software SPCS...
Page 108: ...104 SPCS 200 CareFusion B T P S MicroDL 1 MicroDL 2 CareFusion 3 3 4 SPCS...
Page 109: ...105 MicroDL 1 Fig 1 of enter 3 4 on 2...
Page 110: ...106 MicroDL MicroDL SPCS FEV1 FVC PEF FEV1 FVC SPCS MicroDL ON OFF...
Page 111: ...107 SPCS ENTER 1 ENTER...
Page 112: ...108 MicroDL ENTER ENTER...
Page 113: ...109 SPCS ENTER...
Page 114: ...110 1 1999 ENTER ENTER...
Page 115: ...111 ENTER ENTER ENTER MicroDL...
Page 116: ...112 PEF FEV1 FVC SPCS...
Page 117: ...113 PEF PEF SPCS SPCS PEF FEV1 FVC PEF FEV1 FVC...
Page 119: ...115 ENTER EO E1 E4 ENTER ENTER...
Page 120: ...116 2 ENTER 1 1999 ENTER...
Page 121: ...117 ENTER...
Page 122: ...118 ENTER...
Page 123: ...119 PEF FEV1 FVC FEV1 FVC FEV1FVC SPCS ENTER OFF SPCS FEV1 FVC PEF FEV1 FVC OFF...
Page 126: ...122 1 MicroDL...
Page 128: ...124 FEV1 FVC PEF 2 ENTER 2 MicroDL 2 1 1 ENTER...
Page 131: ...127 3 MicroDL 3 65 65 65 ENTER 2 OFF...
Page 132: ...128 4 MicroDL 4 4 ENTER 1 5 MicroDL 5 5...
Page 135: ...131 MicroDL 10 9 PP3 24 CareFusion B T P S...
Page 136: ...132 MicroDL BTPS 30 3 6 36 SM2125 MicroDL FVC...
Page 137: ...133 Sidex 15...
Page 142: ......
Page 143: ......